Revolution CT

July 14, 2021 Class II

Date
July 14, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Completed
Filed by
GE Healthcare, LLC
Product
Revolution CT
Reason
GE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or smaller than its prescribed value.
Distribution pattern
Worldwide distribution - US Nationwide
Quantity
197 (US = 40; OUS = 157)
Codes and lots
5590000 5590000-6 5590000-2
Recall started
March 19, 2021
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare will perform an initial site assessment and replace the collimator, if needed. This will be followed by bi-weekly remote monitoring of the system s log file. Finally, GE Healthcare will install a currently under development software patch, added to the daily fast calibration to automatically and reliably detect this failure should it occur. All of these actions will be performed at no cost to the customer. For questions call 262-544-3217.
Recall number
Z-2004-2021

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-2004-2021. The agency record governs; check it and the company before acting.