GE Healthcare Optima Mobile X-ray System…

July 25, 2012 Class II

Date
July 25, 2012
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Product
GE Healthcare Optima Mobile X-ray System. The GE Automatic Mobile X-Ray (AMX ) Series are intended to take exposures utilizing film or computed radiography ( CR )
Reason
GE Healthcare has become aware of a software issue on the interface of the mobile radiographic product, Optima XR200amx, ond Optima XR220amx related to Visual indication as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-1-3.
Distribution pattern
Worldwide Distribution (USA) nationwide
Quantity
254 units have been shipped to/installed in the U.S.
Codes and lots
Model Numbers Optima XR200 15kw Optima XR 200 30kw Optima XR 220 15kw Optima XR 220 30kw
Recall started
April 4, 2012
Device type
System, X-Ray, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare will, without charge, remedy the defect or bring the product into compliance with each applicable Federal and IEC standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication to the user or through a GE Healthcare field engineer site visit. A GE Healthcare Service Representative will update the firmware on the system to address the issue. Your proposed corrective action plan (CAP) (Field Modification Instructions 10865) describes the rework plan that GEHC Engineers will make to the imaging system to return it to compliance with the applicable performance standard. It appears to adequately address the problem and is hereby approved.
Recall number
Z-1993-2012

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-1993-2012. The agency record governs; check it and the company before acting.