FDA · Medical device recall
Breast Pack… and 50 more products
- Date
- July 15, 2015
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Completed
- Filed by
- Stradis Healthcare, LLC.
- Reason
- Convenience kits being recalled due to the recall by the manufacturer (Medtronic) of a component (Devon Light Glove) contained in the kits.
- Distribution pattern
- Distribution US nationwide to AL, AK, CA, DE, FL, GA, HI, ID, IL, IN, KS, KY, ME, MD, MA, MI, MS, MO, NM, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA; Canada and Dubai
51 products in this recall
Breast Pack…
- Product
- Breast Pack - contains Devon Light Glove Used during surgery
- Quantity
- 9
- Codes and lots
- Pack 693-196, 693-188, 693-166, 693-112, 888-2583
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-1992-2015
Basic Pack…
- Product
- Basic Pack - contains Devon Light Glove Used during surgery
- Quantity
- 6
- Codes and lots
- Pack 693-203, 888-2375, 888-2396, 888-2430, 888-2495, 888-2594
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-1993-2015
Advanced Vein Care…
- Product
- Advanced Vein Care - contains Devon Light Glove Used during surgery
- Quantity
- Two
- Codes and lots
- Pack 682-956
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-1994-2015
Vein Pack…
- Product
- Vein Pack, Vein Treatment II Pack, Vein Brunswick Pack, Vein No Lidocaine Pack, Vein Ablation Pack - contains Devon Light Glove Used during surgery
- Quantity
- 55
- Codes and lots
- Pack 682-949, 682-937, 682-900, 682-370, 682-218, 682-1264, 682-1529, 682-076, 682-192, 682-962, 682-968, 682-177, 682-394, 682-988, 682-156, 682-063, 682-151, 682-917, 682-250, 682-409, 682-960, 682-945, 682-218, 682-1264, 682-1412, 682-1422
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-1995-2015
Endovenous Kit…
- Product
- Endovenous Kit, Endo Pack and Endovenous Laser Kit - contains Devon Light Glove Used during surgery
- Quantity
- 15
- Codes and lots
- Pack 682-156, 682-1548, 682-931, 682-493, 682-263, 682-931, 682-1419, 682-493, 682-044
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-1996-2015
Phleb Pack and Laser Phleb Pack…
- Product
- Phleb Pack and Laser Phleb Pack - contains Devon Light Glove Used during surgery
- Quantity
- 9
- Codes and lots
- Pack 682-1284, 682-299, 682-410, 682-353, 682-377, 682-548
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-1997-2015
Standard Pack…
- Product
- Standard Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 681-022A
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-1998-2015
Lipo Pack…
- Product
- Lipo Pack, Liposuction Expended Pack, and Liposuction Tray - contains Devon Light Glove Used during surgery
- Quantity
- 13
- Codes and lots
- Pack 681-025, 681-268, 681-254, 681-013, 681-506, 888-2587
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-1999-2015
Lipo/Buttocks Pack…
- Product
- Lipo/Buttocks Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 681-256
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2000-2015
Vaginal Rejuvenation Pack…
- Product
- Vaginal Rejuvenation Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 693-222
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2001-2015
General Pack…
- Product
- General Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 670-115
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2002-2015
Vein & Vascular Pack…
- Product
- Vein & Vascular Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 682-670
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2003-2015
Laser Evlt Pack and Laser One Leg Evlt…
- Product
- Laser Evlt Pack and Laser One Leg Evlt - contains Devon Light Glove Used during surgery
- Quantity
- 3
- Codes and lots
- Pack 682-670, 682-501, 682-547
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2004-2015
Breast Augmentation Pack and Breast Augmentation Expanded Pack…
- Product
- Breast Augmentation Pack and Breast Augmentation Expanded Pack - contains Devon Light Glove Used during surgery
- Quantity
- 24
- Codes and lots
- Pack 681-502, 693-177, 681-243, 681-502, 681-502, 693-158, 682-502, 681-503
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2005-2015
Tummy Tuck Pack…
- Product
- Tummy Tuck Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 693-223
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2006-2015
Surgical Pack and General Surgery Pack…
- Product
- Surgical Pack and General Surgery Pack - contains Devon Light Glove Used during surgery
- Quantity
- 4
- Codes and lots
- Pack 40604EBS, 693-152, 40759MMS, 888-2371
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2007-2015
Mississippi Pack…
- Product
- Mississippi Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 682-1302
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2008-2015
Face Lift Pack…
- Product
- Face Lift Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 693-102
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2009-2015
Plastic Pack and Plastic HNP Pack…
- Product
- Plastic Pack and Plastic HNP Pack - contains Devon Light Glove Used during surgery
- Quantity
- 5
- Codes and lots
- Pack 693-210, 693-054, 693-219, 888-2479, 888-2499
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2010-2015
Suction Pack…
- Product
- Suction Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 670-057
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2011-2015
ACL Pack…
- Product
- ACL Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 693-055
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2012-2015
Hand Pack…
- Product
- Hand Pack - contains Devon Light Glove Used during surgery
- Quantity
- 3
- Codes and lots
- Pack 693-057, 888-2345, 888-2596
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2013-2015
Lima Pack…
- Product
- Lima Pack - contains Devon Light Glove Used during surgery
- Quantity
- 2
- Codes and lots
- Pack 682-927
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2014-2015
Skin and Vein Pack…
- Product
- Skin and Vein Pack - contains Devon Light Glove Used during surgery
- Quantity
- 3
- Codes and lots
- Pack 682-1218
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2015-2015
Vasc Custom Kit…
- Product
- Vasc Custom Kit - contains Devon Light Glove Used during surgery
- Quantity
- 3
- Codes and lots
- Pack 682-126
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2016-2015
Bariatric Pack…
- Product
- Bariatric Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 682-678
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2017-2015
Face Neck Pack…
- Product
- Face Neck Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 693-167
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2018-2015
Abdominal Pack and Abdom Pack…
- Product
- Abdominal Pack and Abdom Pack - contains Devon Light Glove Used during surgery
- Quantity
- 2
- Codes and lots
- Pack 670-056, 681-014
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2019-2015
Stradi 4500…
- Product
- Stradi 4500 - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 682-1102
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2020-2015
Gyn Pack…
- Product
- Gyn Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 693-111
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2021-2015
Head & Neck Pack…
- Product
- Head & Neck Pack - contains Devon Light Glove Used during surgery
- Quantity
- 2
- Codes and lots
- Pack 681-012, 888-2385
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2022-2015
Eye Pack…
- Product
- Eye Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 687-191
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2023-2015
Glaucoma Pack…
- Product
- Glaucoma Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 687-395
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2024-2015
Shoulder Pack…
- Product
- Shoulder Pack - contains Devon Light Glove Used during surgery
- Quantity
- One
- Codes and lots
- Pack 888-2349
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2025-2015
Knee Arthro Pack and Knee Arthro Pack II…
- Product
- Knee Arthro Pack and Knee Arthro Pack II - contains Devon Light Glove Used during surgery
- Quantity
- 3
- Codes and lots
- Pack 888-2359, 888-2531, 888-2351
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2026-2015
ENT Pack…
- Product
- ENT Pack - contains Devon Light Glove Used during surgery
- Quantity
- 2
- Codes and lots
- Pack 888-2352, 888-2434
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2027-2015
Hand Pack…
- Product
- Hand Pack - contains Devon Light Glove Used during surgery
- Quantity
- 4
- Codes and lots
- Pack 888-2376, 888-2397, 888-2414, 888-2471
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2028-2015
Podiatry Pack…
- Product
- Podiatry Pack - contains Devon Light Glove Used during surgery
- Quantity
- 4
- Codes and lots
- Pack 888-2378, 888-2411, 888-2429, 888-2553
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2029-2015
Minor Pack and Minor Pack II Pack…
- Product
- Minor Pack and Minor Pack II Pack - contains Devon Light Glove Used during surgery
- Quantity
- 2
- Codes and lots
- Pack 888-2381, 888-2541
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2030-2015
Neurosurgery Pack…
- Product
- Neurosurgery Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2489
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2031-2015
Cysto Pack…
- Product
- Cysto Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2492
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2032-2015
Open Pack…
- Product
- Open Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2493
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2033-2015
Pediatric Pack…
- Product
- Pediatric Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2494
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2034-2015
Face Lift Pack…
- Product
- Face Lift Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2532
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2035-2015
Set Up Pack…
- Product
- Set Up Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2546
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2036-2015
Blepharoplasty Pack…
- Product
- Blepharoplasty Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2533
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2037-2015
Custom Pack…
- Product
- Custom Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2552
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2038-2015
DND Pack…
- Product
- DND Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2595
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2039-2015
Orthopedic Split Drape Pack…
- Product
- Orthopedic Split Drape Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2604
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2040-2015
Arthroscopy Pack…
- Product
- Arthroscopy Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2605
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2041-2015
Upper Extremity Pack…
- Product
- Upper Extremity Pack - contains Devon Light Glove Used during surgery
- Quantity
- 1
- Codes and lots
- Pack 888-2606
- Recall started
- April 20, 2015
- Device type
- General Surgery Tray
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
- Recall number
- Z-2042-2015
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement reports, recall numbers Z-1992-2015, Z-1993-2015, Z-1994-2015, Z-1995-2015, Z-1996-2015, Z-1997-2015, Z-1998-2015, Z-1999-2015 and others. The agency record governs; check it and the company before acting.