Breast Pack… and 50 more products

July 15, 2015 Class II

Date
July 15, 2015
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Completed
Filed by
Stradis Healthcare, LLC.
Reason
Convenience kits being recalled due to the recall by the manufacturer (Medtronic) of a component (Devon Light Glove) contained in the kits.
Distribution pattern
Distribution US nationwide to AL, AK, CA, DE, FL, GA, HI, ID, IL, IN, KS, KY, ME, MD, MA, MI, MS, MO, NM, NJ, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA; Canada and Dubai

51 products in this recall

Breast Pack…

Product
Breast Pack - contains Devon Light Glove Used during surgery
Quantity
9
Codes and lots
Pack 693-196, 693-188, 693-166, 693-112, 888-2583
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-1992-2015

Basic Pack…

Product
Basic Pack - contains Devon Light Glove Used during surgery
Quantity
6
Codes and lots
Pack 693-203, 888-2375, 888-2396, 888-2430, 888-2495, 888-2594
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-1993-2015

Advanced Vein Care…

Product
Advanced Vein Care - contains Devon Light Glove Used during surgery
Quantity
Two
Codes and lots
Pack 682-956
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-1994-2015

Vein Pack…

Product
Vein Pack, Vein Treatment II Pack, Vein Brunswick Pack, Vein No Lidocaine Pack, Vein Ablation Pack - contains Devon Light Glove Used during surgery
Quantity
55
Codes and lots
Pack 682-949, 682-937, 682-900, 682-370, 682-218, 682-1264, 682-1529, 682-076, 682-192, 682-962, 682-968, 682-177, 682-394, 682-988, 682-156, 682-063, 682-151, 682-917, 682-250, 682-409, 682-960, 682-945, 682-218, 682-1264, 682-1412, 682-1422
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-1995-2015

Endovenous Kit…

Product
Endovenous Kit, Endo Pack and Endovenous Laser Kit - contains Devon Light Glove Used during surgery
Quantity
15
Codes and lots
Pack 682-156, 682-1548, 682-931, 682-493, 682-263, 682-931, 682-1419, 682-493, 682-044
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-1996-2015

Phleb Pack and Laser Phleb Pack…

Product
Phleb Pack and Laser Phleb Pack - contains Devon Light Glove Used during surgery
Quantity
9
Codes and lots
Pack 682-1284, 682-299, 682-410, 682-353, 682-377, 682-548
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-1997-2015

Standard Pack…

Product
Standard Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 681-022A
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-1998-2015

Lipo Pack…

Product
Lipo Pack, Liposuction Expended Pack, and Liposuction Tray - contains Devon Light Glove Used during surgery
Quantity
13
Codes and lots
Pack 681-025, 681-268, 681-254, 681-013, 681-506, 888-2587
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-1999-2015

Lipo/Buttocks Pack…

Product
Lipo/Buttocks Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 681-256
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2000-2015

Vaginal Rejuvenation Pack…

Product
Vaginal Rejuvenation Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 693-222
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2001-2015

General Pack…

Product
General Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 670-115
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2002-2015

Vein & Vascular Pack…

Product
Vein & Vascular Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 682-670
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2003-2015

Laser Evlt Pack and Laser One Leg Evlt…

Product
Laser Evlt Pack and Laser One Leg Evlt - contains Devon Light Glove Used during surgery
Quantity
3
Codes and lots
Pack 682-670, 682-501, 682-547
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2004-2015

Breast Augmentation Pack and Breast Augmentation Expanded Pack…

Product
Breast Augmentation Pack and Breast Augmentation Expanded Pack - contains Devon Light Glove Used during surgery
Quantity
24
Codes and lots
Pack 681-502, 693-177, 681-243, 681-502, 681-502, 693-158, 682-502, 681-503
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2005-2015

Tummy Tuck Pack…

Product
Tummy Tuck Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 693-223
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2006-2015

Surgical Pack and General Surgery Pack…

Product
Surgical Pack and General Surgery Pack - contains Devon Light Glove Used during surgery
Quantity
4
Codes and lots
Pack 40604EBS, 693-152, 40759MMS, 888-2371
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2007-2015

Mississippi Pack…

Product
Mississippi Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 682-1302
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2008-2015

Face Lift Pack…

Product
Face Lift Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 693-102
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2009-2015

Plastic Pack and Plastic HNP Pack…

Product
Plastic Pack and Plastic HNP Pack - contains Devon Light Glove Used during surgery
Quantity
5
Codes and lots
Pack 693-210, 693-054, 693-219, 888-2479, 888-2499
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2010-2015

Suction Pack…

Product
Suction Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 670-057
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2011-2015

ACL Pack…

Product
ACL Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 693-055
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2012-2015

Hand Pack…

Product
Hand Pack - contains Devon Light Glove Used during surgery
Quantity
3
Codes and lots
Pack 693-057, 888-2345, 888-2596
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2013-2015

Lima Pack…

Product
Lima Pack - contains Devon Light Glove Used during surgery
Quantity
2
Codes and lots
Pack 682-927
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2014-2015

Skin and Vein Pack…

Product
Skin and Vein Pack - contains Devon Light Glove Used during surgery
Quantity
3
Codes and lots
Pack 682-1218
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2015-2015

Vasc Custom Kit…

Product
Vasc Custom Kit - contains Devon Light Glove Used during surgery
Quantity
3
Codes and lots
Pack 682-126
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2016-2015

Bariatric Pack…

Product
Bariatric Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 682-678
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2017-2015

Face Neck Pack…

Product
Face Neck Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 693-167
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2018-2015

Abdominal Pack and Abdom Pack…

Product
Abdominal Pack and Abdom Pack - contains Devon Light Glove Used during surgery
Quantity
2
Codes and lots
Pack 670-056, 681-014
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2019-2015

Stradi 4500…

Product
Stradi 4500 - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 682-1102
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2020-2015

Gyn Pack…

Product
Gyn Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 693-111
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2021-2015

Head & Neck Pack…

Product
Head & Neck Pack - contains Devon Light Glove Used during surgery
Quantity
2
Codes and lots
Pack 681-012, 888-2385
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2022-2015

Eye Pack…

Product
Eye Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 687-191
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2023-2015

Glaucoma Pack…

Product
Glaucoma Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 687-395
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2024-2015

Shoulder Pack…

Product
Shoulder Pack - contains Devon Light Glove Used during surgery
Quantity
One
Codes and lots
Pack 888-2349
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2025-2015

Knee Arthro Pack and Knee Arthro Pack II…

Product
Knee Arthro Pack and Knee Arthro Pack II - contains Devon Light Glove Used during surgery
Quantity
3
Codes and lots
Pack 888-2359, 888-2531, 888-2351
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2026-2015

ENT Pack…

Product
ENT Pack - contains Devon Light Glove Used during surgery
Quantity
2
Codes and lots
Pack 888-2352, 888-2434
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2027-2015

Hand Pack…

Product
Hand Pack - contains Devon Light Glove Used during surgery
Quantity
4
Codes and lots
Pack 888-2376, 888-2397, 888-2414, 888-2471
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2028-2015

Podiatry Pack…

Product
Podiatry Pack - contains Devon Light Glove Used during surgery
Quantity
4
Codes and lots
Pack 888-2378, 888-2411, 888-2429, 888-2553
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2029-2015

Minor Pack and Minor Pack II Pack…

Product
Minor Pack and Minor Pack II Pack - contains Devon Light Glove Used during surgery
Quantity
2
Codes and lots
Pack 888-2381, 888-2541
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2030-2015

Neurosurgery Pack…

Product
Neurosurgery Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2489
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2031-2015

Cysto Pack…

Product
Cysto Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2492
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2032-2015

Open Pack…

Product
Open Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2493
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2033-2015

Pediatric Pack…

Product
Pediatric Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2494
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2034-2015

Face Lift Pack…

Product
Face Lift Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2532
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2035-2015

Set Up Pack…

Product
Set Up Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2546
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2036-2015

Blepharoplasty Pack…

Product
Blepharoplasty Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2533
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2037-2015

Custom Pack…

Product
Custom Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2552
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2038-2015

DND Pack…

Product
DND Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2595
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2039-2015

Orthopedic Split Drape Pack…

Product
Orthopedic Split Drape Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2604
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2040-2015

Arthroscopy Pack…

Product
Arthroscopy Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2605
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2041-2015

Upper Extremity Pack…

Product
Upper Extremity Pack - contains Devon Light Glove Used during surgery
Quantity
1
Codes and lots
Pack 888-2606
Recall started
April 20, 2015
Device type
General Surgery Tray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Medtronic sent an Urgent Medical Device Recall letter dated April 16, 2015, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to quarantine and discontinue use of the affected device. Complete the Recalled Product Return Form and fax to 800-895-6140 or email to feedback.customerservice@Covidien.com. Customers with questions regarding return of the product should call 1-727-642--4124.
Recall number
Z-2042-2015

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-1992-2015, Z-1993-2015, Z-1994-2015, Z-1995-2015, Z-1996-2015, Z-1997-2015, Z-1998-2015, Z-1999-2015 and others. The agency record governs; check it and the company before acting.

Report a mistake on this page