FDA · Medical device recall

GE Healthcare… and 1 more product

July 9, 2014 Class II

Date
July 9, 2014
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Reason
GE Healthcare has become aware of a potential safety issue due to a collimator installation error during a service maintenance activity involving the Proteus XR/a and Revolution XR/d X-ray imaging systems. . There was a reported incident of a collimator fall which caused a serious patient injury.
Distribution pattern
Worldwide Distribution - USA including PR. DC, GUAM, all states except VT. Internationally to the following countries: ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BENIN, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BULGARIA, BURKINA FASO, CANADA, CHILE, CHINA, COLOMBIA, CZECH REPUBLIC, DOMINICAN REPUBLIC, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, FRENCH POLYNESIA, GERMANY, GREECE, GUATEMALA, HONDURAS, HONG KONG, HUNGRY, ICELAND, INDONESIA, IRAQ, IRELAND, ISRAEL, ITALY, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LIBYAN ARAB JAMHIRIYA, LITHUANIA, MALAYSIA, MALTA, MARTINIQUE, MEXICO, MOROCCO, NETHERLANDS, NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, PAKISTAN, PANAMA, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SUDAN, SWEDEN, SWITZERLAND, TAIWAN, TANZANIA, THAILAND, TUNISIA, TURKEY, UGANDA, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VENEZUELA, VIETNAM YEMEN.

2 products in this recall

GE Healthcare…

Product
GE Healthcare, Proteus XR/a. Proteus XR/a Intended Use: Is intended for use in generating radiographic images of human anatomy in all general purpose diagnostic procedures. This device is not intended for mammographic applications.
Quantity
3460
Codes and lots
Mfg Lot or Serial # System ID 00000023354HL4 804289HDFER 00000017868HL1 801268ER 00000024445HL9 804747DHPER 00000018247HL7 972596PER3 00000021176HL3 804541ER 00000018523HL1 408977ER3 00000019915HL8 772468ER1 00000018898HL7 813571ERPRO 00000018576HL9 850747GCER 00000024536HL5 904391RAD1 00000019171HL8 561395RAD 00000020749HL8 773282TOMO8 00000034015HL8 847316PROER1 00000012567HL4 502868RAD2 00000017199HL1 972596OPRAD1 00000017648HL7 405271UPX1 00000032656HL1 407518PR1 00000041533HL1 580249SPRO2 00000022569HL8 614868PRO4 00000012718HL3 706257SFP 00000020559HL1 770748PER 00000033658HL6 704783PRO1 00000019846HL5 423493ER 00000031607HL5 605342RAD 00000012703HL5 915573RAD1 00000012705HL0 915573RAD2 00000020034HL5 410392R3 00000022299HL2 706554MSR1 00000003188HL2 901722RAD1 00000019209HL6 718780RAD1 00000011073HL4 865425RAD2 00000018254HL3 281338RAD 00000020919HL7 801292RM3 00000041379HL9 580249SPRO 00000038277HL0 408885SVC 00000029264HL9 561863RM6 00000038416HL4 702382XR1 00000030772HL8 201291PROTEUS 00000018574HL4 201358ER1 00000018682HL5 201358ER2 00000023348HL6 201418RAD1 00000021851HL1 201418RAD3 00000030395HL8 201599PRO 00000100468HL8 201795ER 00000032859HL1 202269PRO 00000013889HL1 202715PRO 00000016974HL8 203679RAD 00000078697HL0 203688PROTLW2 00000021322HL3 203797PROT1 00000022725HL6 203797PROT2 00000028953HL8 203797PROT3 00000027089HL2 203852ED1 00000030390HL9 203852RM1 00000028277HL2 203852RM5 00000023570HL5 205387PRO 00000044402HL6 205750RAD 00000013259HL7 206215RM1 00000037711HL9 206248RAD1 00000034679HL1 206302XRM1 00000013114HL4 206543HECED 00000016238HL8 207454CRAD1 00000080456HL7 207664PROTEUS 00000039888HL3 207795CM3 00000018794HL8 208227RM1 00000019521HL4 208227RM5 00000040501HL9 208265PROT 00000020176HL4 208359RM2 00000020667HL2 208373RAD2 00000023516HL8 208373RAD3 00000019212HL0 208381ED1 00000103408HL1 208402RAD1 00000042856HL5 208455RAD3 00000041973HL9 208455RAD4 00000035259HL1 208463RAD1 00000015447HL6 208463RAD2 00000029208HL6 208476CRAD 00000037477HL7 208524ROOM2 00000013648HL1 208529ER 00000038278HL8 208667KS 00000029103HL9 208746VX1A 00000017660HL2 208756SX2 00000014152HL3 208785RAD 00000014711HL6 208882GX1A 00000010593HL2 208893RAD2 00000024418HL6 208983SRAD 00000043582HL6 209223PROTAM1 00000023095HL3 209385RAD1 00000024355HL0 209524GR5 00000022811HL4 209577RAD3 00000047950HL1 209578ER1 00000059852HL4 209664RAD3 00000043977HL8 209754RAD1 00000029481HL9 209832RAD1 00000047492HL4 209835RAD2 00000020668HL0 209848OVRAD2 00000014446HL9 209943SJRM2 00000029151HL8 210297BMCR5 00000039510HL3 210297NCER1 00000025154HL6 210297NCRAD 00000025532HL3 210297NERAD 00000029153HL4 210297SEBR4 00000046415HL6 210575LCRAD 00000045916HL4 210590OWRAD 00000031131HL6 210614PRO 00000032826HL0 210614SPRO 00000029511HL3 210616MCTRAD2 00000041870HL7 210616PRO 00000029671HL5 210616R2 00000044731HL8 210617NEI1 00000029265HL6 210646NEMR1 00000037581HL6 210646RAD2 00000016636HL3 210704SRER 00000032255HL2 210705MCER 00000042165HL1 210921PRO 00000023092HL0 212241RAD 00000041980HL4 212312PRO1 00000042075HL2 212312PRO2 00000019847HL3 212570RAD 00000021798HL4 212604PROT1 00000021682HL0 212604PROT3 00000019515HL6 213744PROTXR2 00000070214HL2 214266IRVPRO 00000036459HL6 214266RAD1 00000013453HL6 214320PROT2 00000016419HL4 214320PTOMO 00000031876HL6 214345RAD1 00000094194HL8 214491RPG 00000036568HL4 215441PROT1 00000024635HL5 215453PRO3 00000043808HL5 215481PROT1 00000014447HL7 215918RAD1 00000014709HL0 215918RAD2 00000016795HL7 215952PROT 00000043807HL7 215955PROER 00000045769HL7 215955PROER2 00000014147HL3 216343R3 00000014298HL4 216343R4 00000020177HL2 216363RAD 00000044378HL8 216363RAD1 00000047424HL7 216778RAD1 00000036104HL8 216844PRO1 00000024694HL2 217222QR6 00000012561HL7 217258SR3 00000024632HL2 217333LXR2 00000037884HL4 217333MXR1 00000028272HL3 217337CXR3 00000081388HL1 217366CWXR2 00000021371HL0 217366CXR1 00000034341HL8 217383CER1 00000034433HL3 217383CER3 00000034017HL4 217383CXR20 00000033378HL1 217383CXR24 00000045391HL0 217431CFAXR1 00000026046HL3 217431CXR1 00000035175HL9 217443UXR7 00000024771HL8 217732LPR1 00000017441HL7 217788PRT1 00000032373HL3 217824PRT 00000016249HL5 218759RAD 00000102146HL8 218935PRO 00000019217HL9 218935RAD 00000017658HL6 219284ER 00000014036HL8 219284PRO1 00000014449HL3 219284RM3N 00000016970HL6 219347RAD 00000017866HL5 219492RAD1 00000018575HL1 219492RAD2 00000022813HL0 219757ER 00000028862HL1 219757WP1 00000014713HL2 225638R1 00000028856HL3 225761R5 00000033648HL7 228436XR1 00000065125HL7 228818PROT4 00000041120HL7 229243RAD1 00000023084HL7 229249SSC 00000027088HL4 229259OUT 00000098879HL0 229276CR3 00000035913HL3 229276PROT 00000019914HL1 229890CR4 00000050612HL1 231487ED 00000019168HC4 231547PROTEUS 00000015117HL5 231726PROTEUS 00000028660HL9 231924ER 00000016260HL2 231924RM3 00000028859HL7 239254RAD 00000033657HL8 239368XR1 00000050086HL8 239403RAD 00000025533HL1 239432XR1 00000022345HL3 240379RAD 00000028086HL7 248601RAD 00000036859HL7 248615PROT 00000035349HL0 248652ER 00000014155HL6 248652PROT3 00000025530HL7 248652RAD4 00000018536HL3 248858ER 00000032480HL6 248858ER1 00000018139HL6 248937RM3 00000020846HL2 251275RAD 00000035138HL7 251368ARAD 00000017442HL5 251435I02 00000015449HL2 251435R1 00000023383HL3 252209RCR3 00000035857HL2 252436RAD1 00000035561HL0 252436RAD2 00000036449HL7 252443NGER1 00000036447HL1 252443NGER2 00000023350HL2 252482CR2 00000019842HL4 252522LRAD1 00000036148HL5 252637ECR2 00000035260HL9 252757ORTHO1 00000036000HL8 252757ORTHO3 00000023877HL4 252794BR2 00000035031HL4 253597XRM1 00000098068HL0 253680PROTEUS 00000035557HL8 253735MS1 00000030194EL1 253736RANIER 00000077534HL6 253841DINW1 00000021789HL3 254562RAD2 00000016650HL4 254897GRRAD 00000017859HL0 256235ORT1 00000028657HL5 256974PR1 00000039692HL9 260373RM1 00000028017HL2 262249PROT 00000015842HL8 262268RAD1 00000014451HL9 262321RAD1 00000020329HL9 262334RAD1 00000034554HL6 262363PROT 00000017443HL3 262387RAD1 00000003688HL0 262642PROT 00000088865HL1 262687PRO1 00000025413HL6 262741PR2 00000022563HL1 262767PRO 00000051000HL8 262785PRO1 00000016959HL9 262785RM1 00000047494HL0 262884PRO1 00000032114HL1 262928ER1 00000031765HL1 262928ER2 00000032484HL8 269341PROTEUS 00000041974HL4 269373PROTEUS 00000046319HL0 269463PROTEUS 00000024173HL7 269637PROTEUS 00000033227HL0 269649PROTEUS 00000025412HL8 269945PROTEUS 00000047036HL9 269968PROTEUS 00000015852HL7 270259R3 00000030473HL3 270444LR4 00000030471HL7 270444MNOUT 00000030068HL1 270444MNPOC 00000023523HL4 270745RAD1 00000028151HL9 270745TOMO 00000017867HL3 276546PRORAD 00000023093HL8 276964PRORAD1 00000022807HL2 276964RAD 00000047497HL3 281348PROTEUSER 00000038319HL0 281364PROD 00000015441HL9 281392PRO 00000089284HL4 281395KATYPRO 00000017247HL8 281558PRO2 00000029105HL4 281599PRO 00000032283HL4 281837BPROD1 00000072827HL9 281923PWPRO1 00000046867HL8 281929PRORAD 00000104061HL7 281FCLPPROT 00000021313HL2 301290RAD1 00000042895HL3 301609ER 00000022488HL1 301652RAD 00000025059HL7 301698ER2 00000024710HL6 301698RM3 00000021316HL5 301723RAD4 00000041539HL8 301762RAD1 00000036176HL6 301868R 00000017433HL4 301896RAD1 00000017858HL2 302392PRO 00000029480HL1 302421PRO 00000037479HL3 3024307 00000046647HL4 302430RAD 00000020922HL1 302651PRO3 00000021002HL1 302651PRO4 00000042767HL4 302744ER 00000038053HL5 302744RAD 00000023155HL5 303320ER1 00000039726HL5 303338RAD 00000040767HL6 303426ER 00000023152HL2 303426RM2 00000016425HL1 303450ED 00000046870HL2 303563XR 00000031012HL8 303629RM05 00000023146HL4 303629RM06 00000031668HL7 303629RM08 00000028860HL5 303661RAD 00000022352HL9 303673TXR1 00000023028HL4 303673TXR2 00000035896HL0 303695RAD1 00000020916HL3 303699ER 00000021680HL4 303730RM5 00000030267HL9 303778EDXR 00000014440HL2 304257PROT 00000078974HL3 304258RAD1 00000015115HL9 304347XR4 00000011346HL4 304354PROD 00000018408HL5 304424PRO 00000039595HL4 304424RM6 00000018250HL1 304424RM8 00000032376HL6 304598PROT1 00000020043HL6 304599PRODA 00000033374HL0 304624PRT 00000032212HL3 304792PROT1 00000044436HL4 304927R2 00000023082HL1 305295R1 00000041178HL5 305325RM9 00000034576HL9 305345STRAD 00000031033HL4 305682ER1 00000036548HL6 305823ER2 00000016237HL0 307265GERAD 00000042835HL9 307382RAD1 00000034081HL0 307527PROTEUS 00000040966HL4 307634XRA 00000042184HL2 308534NPOXR 00000029991HL7 308928RAD 00000035352HL4 309284XR1 00000015445HL0 309343CR2 00000028486HL9 309454BER 00000058231HL2 309454BX3A 00000034154HL5 309582PROT 00000015195HL1 309655FNCR1 00000015199HL3 309655FNCR2 00000016972HL2 309655FXR6 00000018796HL3 309655MORTON2 00000031034HL2 309662JER 00000029484HL3 309663MHXR1 00000026037HL2 309664CBN2 00000016427HL7 309944HHPROT 00000015856HL8 313359PROT 00000012556HL7 313369PROT 00000036738HL3 313499PROT 00000029439HL7 313640PROT1 00000038206HL9 314251JDRM2 00000045768HL9 314251SJBL 00000045767HL1 314251SJBR 00000037336HL5 314251SJMXR1 00000030698HL5 314344DR6 00000027922HL4 314577CP1 00000019912HL5 314625WER 00000013887HL5 314768MRM8 00000041965HL5 314879CC 00000080625HL7 314919JRAD1 00000036308HL5 315332NR4 00000060161HL7 315332RM3 00000044729HL2 315338ER 00000016639HL7 315782RAD1 00000016642HL1 315782RAD2 00000091297HL2 315787TB1 00000017204HL9 316689WC3 00000017188HL4 316689WC4 00000016430HL1 316722WWRAD 00000023876HL6 316945PRO 00000042858HL1 316962PRO 00000037925HL5 317415PROT 00000034274HL1 317621PROT 00000019848HL1 317745HEPR5 00000019375HL5 317745HEPR6 00000028861HL3 317888PRO1 00000039277HL9 317888PRO2 00000041057HL1 318251PRO 00000029210HL2 318254PRO 00000047207HL6 318255PRO 00000033935HL8 318335ORM1 00000011582HL4 318425LR1 00000029482HL7 318445LCXR1 00000086525HL3 318769ERM1 00000034777HL3 318878PRO 00000032377HL4 319235PROT 00000032862HL5 319268PROT 00000025362HL5 319268SER 00000035914HL1 319352PROT 00000015850HL1 319376PROT 00000010592HL4 319382PROTXR 00000036858HL9 319642PROT 00000029666HL5 319863PROT 00000017448HL2 320632RAD 00000017446HL6 320654R1 00000029928HL9 325670PRO2 00000029941HL2 325670PRO3 00000016796HL5 3271539RAD 00000016751HL0 330332R4 00000043314HL4 330386ER 00000029097HL3 3304382 00000015442HL7 330438R3 00000014306HL5 330489XR3 00000041538HL0 330543XRA 00000040906HL0 330834RAD 00000108822HL8 331551DMGLPROT3 00000023519HL2 332047 00000047158HL1 333PRO 00000046877HL7 334244RAD2 00000022342HL0 334636TRM1 00000033481HL3 336246AMRAD1 00000046734HL0 336719NR140 00000045750HL7 336719NR144 00000010846HL4 336768OSCD 00000010879HL5 336768OSEF 00000010845HL6 336768OSGH 00000022805HL6 336768OSR1 00000046645HL8 337355RAD1 00000038339HL8 337457RAD1 00000034557HL9 337981ER1 00000029190HL6 337981R3 00000033416HL9 337981RAD3 00000080459HL1 33RM4 00000040585HL2 352253RAD1 00000036373HL9 352394RM3 00000023878HL2 352401RAD2 00000015124HL1 352401RAD3 00000029206HL0 352596ERPRO 00000099463HL2 352596PROM 00000045275HL5 352597RAD1 00000032481HL4 360412XR1 00000043297HL1 360414OTHO 00000023757HL8 360414SJOP1 00000012540HL1 360414SJRM4 00000021172HL2 360457ORTHO 00000036280HL6 360493PROT 00000016973HL0 360566RM1 00000028479HL4 360651XR1 00000003535HL2 360754XRM2 00000043001HL7 360992CLK1 00000047153HL2 361275RM1 00000025528HL1 361572CER 00000042524HL9 361698DPR1 00000032653HL8 361857DOR1 00000022340HL4 361980RAP 00000052253HL2 373ORTHO 00000023521HL8 386238FHR1 0000002442OHL2 386238FHR3 00000101039HL6 386326ER 00000028014HL9 386446RM1 00000028855HL5 386446RM2 00000035556HL0 386774RM1 00000034556HL1 386774RM2 00000041537HL2 386FHIRAD 00000043002HL5 386HPORAD 00000032169HL5 386PORAD 00000031204HL1 386TWLRAD 00000016631HL4 401274ER1 00000029101HL3 401431PR1 00000020041HL0 401727P1 00000034920HL9 401766ARP1 00000021678HL8 401782SR1 00000022544HL1 401782SR2 00000029799HL4 401782SR3 00000094369HL6 402393MERAD 00000016794HL0 402481BX3 00000016244HL6 402481LGMX10 00000013775HL2 402481PLRAD 00000032116HL6 402643SMX1 00000020167HL3 402685RAD 00000052993HL3 402745PROT 00000039202HL7 402768RAD1 00000024449HL1 402925RAD 00000030157HL2 402947WRAD 00000037710HL1 404265ER1 00000046411HL5 404265PROT 00000034680HL9 404321PRO1…
Recall started
June 4, 2014
Device type
System, X-Ray, Stationary
Root cause (FDA)
Equipment maintenance
What the company told customers
GE Healthcare sent an Urgent Medical Device Correction letter dated June 4, 2014, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions in regard to this recall the consignees in the United States were advised to call 800-437-1171. For other countries contact your local GE Healthcare Service Representative. For questions regarding this recall call 800-437-1171.
Recall number
Z-1988-2014

GE Healthcare…

Product
GE Healthcare, Revolution XR/d. Revolution XR/d Intended Use: The Revolution XR/d Digital Radiographic imaging system is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammographic applications.
Quantity
1053
Codes and lots
Mfg Lot or Serial # System ID 00000000001837 815338WXRD 00000000002821 815759RAD2 00000000003407 205206SWXRD1 00000000001153 205838AXRD1 00000000003535 205939SVER 00000000002574 209669XRD 00000000001745 215481DR19 00000000001320 215481DR21 00000000001315 215481DR2BLAIR 00000000002962 215481XRDER1 00000000003098 251435XRD1 00000000003049 251435XRD2 00000000001366 406496XRD 00000000001380 413447XRD 00000000002377 417347ED1 00000000002584 417451NMC 00000000002257 503GICXRD 00000000002243 505834XRD 00000000002458 516IPKXRD 00000000003171 567661XRD 00000000002865 601443XRD1 00000000002007 606549XRD 00000000002590 618993MDRF 00000000002378 630236XRD 00000000002722 630275ERXRD 00000000002480 630856ER1 00000000003347 708783R8 00000000003524 757886ER 00000000001026 773MTSXRD1 00000000003954 814868RAD 00000000003465 815740XRDER 00000000002414 815933XRD2 00000000002501 847437GEXRD3 00000000003112 850474XRD 00000000002839 850474XRD2 00000000002054 928669XRD 00000000002688 985230RAD4 00000000002869 CLAKEXRD 00000000002488 EBLUFFXRD 00000000004228 505784XRD 00000000004402 843390CON 00000000007854 864560IP3R 00000000003119 210292WHORT 00000000001454 303398XRD 00000000001050 312666VAHUM 00000000001148 312CCHER2 00000000001199 312CCHRAD4 00000000001156 312CCHRAD5 00000000001176 312CCHRAD6 00000000001230 312CCHRAD7 00000000001220 312CCHRAD8 00000000001464 314525SARM8 00000000001193 404321XRD 00000000001357 509942XRD 00000000001087 510372XRD 00000000002395 601264XRD1 00000000002086 605322XRD 00000000001792 605719XRD1 00000000001534 617726XR8Y3 00000000003984 631SLOANXRD 00001126 650493XRD 00000000001091 775786DR01 00000000001760 775885XRD 00000000001075 865545DR2 00000000001368 916366XRD 00000000001089 973971PED1 00000000002124 JAC1D33 00000000001323 JAC1D40 00000000001871 JAC3D13 00000000002239 JAC3E24 00000000002227 JAC3E48 00000000001887 JAC3E53 00000000004024 727893ER1 00000000001274 281338XDER2 00000000001940 509474XRD 00000000002001 415600XRD1 00000000001802 865481DR1 00000000001660 972566XRD1 00000000002675 601288XRD1 00000000001136 281338XD1 00000000001547 678945XRD 00000000001997 864560ER1 00000000001448 2027825A26 00000000003595 203694IC46 00000000001396 205592XRD1 00000000001218 205783PDR2 00000000001458 208737XRD 00000000002481 210646LXRD 00000000001795 212639XRD2 00000000002205 212639XRD3 00000000002560 217258SBER 00000000001666 217876DR8 00000000002296 225358XRD 00000000001621 229312MPXRD 00000000001643 229312MPXRD2 00000000001462 253426XRD1 00000000002000 253426XRD2 00000000022392 256265XR2 00000000002120 269983XRD 00000000002678 303932XRD 00000000001784 310319XRD 00000000001444 310319XRD2 00000000002949 310319XRD3 00000000003542 314251JDRM4 00000000002158 314525AIXRD1 00000000003356 314525AIXRD2 00000000001682 314525SARM5 00000000001580 314569SJCCX1 00000000007286 315449RAD2 00000000001683 317216XRD 0000000000AGA6 318456BRM2 00000000002661 330841XRD7 00000000001849 334293RADER 00000000001720 336713BCC1 00000000001952 336713BCC2 00000000001957 336713BCC4 00000000002125 336716BCR7 00000000001714 404686CRAD8 00000000002404 405271XRD 00000000002470 408236KSXRD2 00000000002564 408236KSXRD3 00000000002363 408236KSXRD7 00000000001318 409772ER2 00000000001853 409899RAD2 00000000003065 410605XRD1 0000FW21530260 412359XRD 00000000001723 415221XRD1 00000000002282 415353XRD1 00000000002502 419483DIGITAL1 00000000001247 423224DR2 00000000001876 423586DR1 00000000002588 503413LEXRD 00000000001874 504447XRD 00000000002462 507437XRD 00000000003045 509946XRD 00000000001375 515226XRD 00000000001413 515241XRD 00000000001710 515241XRD2 00000000002326 516562ERXRD1 00000000002307 518243RM7 00000000001570 518292XRD2 00000000001575 518292XRD3 00000000002427 520574XRD 00000000002889 520TMCXRD 0000000000A885 540332RM1 00000000001200 540332RM5 00000000003258 559325XRD 00000000002793 559459XRD2 00000000002373 570523XRD1 00000000001877 574523ER1 00000000002340 601426XRD1 00000000001057 602239ER2 00000000001635 610770XRD 00000000001178 610954XRD 00000000001830 615327XRD 00000000002800 615338RAD 00000000001831 617340XRD 00000000001675 617726XR1Y6 00000000001204 617726XR3Y3 00000000002189 617726XR5Y3 00000000001834 617726XR7Y3 00000000003920 617732XRD12 00000000001844 617732XRD14 00000000001093 630978XRDE185 00000000001121 631968ERDIG 00000000002099 650497CHXRD1 00000000001868 650497XRD1 00000000002765 702384S05 00000000001326 702855XRD 00000000001211 703696XRD 00000000002328 707442HUMXRD1 00000000001358 708216XRD2 00000000002638 708342XRD 00000000002344 708799XRD 00000000001623 712852XRD 00000000001293 713394ERXRD1 00000000002538 713790MMXRD1 00000000001917 713794ADRA4 00000000001118 713873BTXRD11 00000000001260 715361XRD 00000000001684 718226NORTHER 00000000002871 718240XRD 00000000002311 718901XRD 00000000002520 718963XRD 00000000001271 719867RM11 00000000001442 720225ER 00000000001117 727298ER2 00000000001629 727733RM1 00000000001119 734712CSB1 00000000001352 734712ER1 00000000002818 734712ER2 00000000001130 757594RM1 00000000001059 757764RAD1 00000000001250 770888XRD 00000000001776 803779R1 00000000003150 804594RM1 00000000002720 804MRMCER1 00000000002917 812858GWER1 00000000004401 814437XRD1 00000000002528 814437XRD2 00000000001881 814452XRD1 00000000002017 814452XRD2 00000000002537 815933XRD 00000000001518 816276XRD 00000000001312 831625XRD 00000000002365 843477VAXRD 00000000002857 843573WAXRD 00000000002621 843832ER 00000000001401 843832XRD3 00000000001801 847490XRD3 00000000001707 847618CSMXRD 00000000001730 847723XRD3 00000000001992 847723XRDER 00000000007727 847913DRAD1 00000000001341 847985SICXRD 00000000001051 847990ER1 00000000001264 847STAERXRD 00000000001471 864560BJWR2 00000000002037 864560ER2 00000000002030 864560ER3 00000000001843 865544EDXRD 00000000001842 865544OPXRD1 00000000001751 865545DR1 00000000002509 865977DR 00000000001996 901380FCI 00000000002200 903614CSMERXRD 00000000002270 904202PD1RM1A 00000000003274 909464ER 00000000001847 910822VAXRD 00000000001287 912350MMCER2 00000000001479 928779RM2 00000000001739 941629XRD 00000000002199 970544RM1 00000000001480 972566XRDER1 00000000007833 972731RAD1 00000000002840 973579ER 00000000001916 973579RM3 00000000007245 973971PED3 00000000001880 973971RAD1 00000000002407 JMMCXRD2 00000000001965 901516XRD 00000000001405 215513XRD 00000000001237 918494D8 00000000001840 615338XRD 00000000002735 801587DX2 00000000003803 801587DX1 00000000007894 864560IP2R 00000000001216 602239RMG 00000000001022 919684A3DIG 00000000002984 408236KSXRD8 00000000002540 562424XRD 00000000001343 205783PDR1 00000000001951 205996MCXRD 00000000001308 212639XRD1 00000000001516 215707CHEST 00000000001077 216844XRD2 00000000002475 256265XR3 00000000001850 270444MNXRD 00000000001289 315449CHEST 00000000002850 330841XRD9 00000000002549 337981XRD 00000000001772 401737KREV 00000000001386 402219XRD1 00000000001361 402219XRD2 00000000001733 412777XRD2 00000000002281 415221XRD2 00000000002335 415750XRD1 00000000002870 478757ER 00000000003067 505599XRD 00000000001851 509473DR2 00000000007747 608263XRD1 00000000001067 608263XRD2 00000000001778 610526XRD1 00000000002700 630856QXI 00000000003703 630978DEF 00000000004169 650497SHXRD1 00000000001697 650497XRD2 00000000002923 760241XRD 00000000007239 773967XRD2 00000000001223 803771CHT 00000000003697 814437XRD3 00000000003430 910215XRDC 00000000003377 952746XRD1 00000000002847 954987PED1 00000000001886 985898ER1 00000000001808 205295WXRD 00000000003285 215710DRPAT 00000000002466 415668XRD2 00000000003502 305662RM2 00000000003434 609267DEF 00000000003426 773STJXRD317 00000000002353 205939SVXRD 00000000002399 217223XRD 00000000002385 256265XR1 00000000001968 310825XRD2 00000000003353 310825XRD3 00000000002419 330729XRD 00000000002642 360493DR2 00000000002559 360493DR3 00000000002151 402449CXRD1 00000000002005 412942XRD 00000000002171 419483DIGITAL2 00000000002572 716250RAD1 00000000002367 912920XRD 00000000002424 216844PEDR2 00000000003300 541667XRD 00000000002396 570969XRD 00000000002997 609835XRD 00000000002928 636947XRDER 00000000002897 763689XRD 00000000002972 865425XRD1 00000000001461 949364ER 00000000003351 205554VAXRD1 00000000002894 213485XRD 00000000002586 228818OSXRD 00000000002766 253952XRD1 00000000001186 281420XRDL 00000000003818 281943DRAD 00000000002977 301784FCI 00000000002959 303320XRD1 00000000003097 304379XRD 00000000004093 305662ORTHO1 00000000001749 310319XRDER1 00000000003268 313343XRD1 00000000002522 313593XRD1 00000000003497 314251JDHC1 00000000003513 314251JDHC2 00000000003184 318335XRD 00000000002995 318675ACCR1 00000000002447 318675ACCR2 00000000002809 330489XRD 00000000002472 336716BCR1 00000000002780 336716BCR3 00000000002676 337261UXRD1 00000000003930 337261UXRD2 00000000002887 402336XRD 00000000003975 432685DDR3 00000000003313 503215BRXRD 00000000003319 503215ERXRD 00000000002948 509474XRDTR1 00000000002828 540776XRD4 00000000003966 541677MDR1 00000000003962 541677MDR2 00000000002655 570547FCIM 00000000003225 573884XRAY 00000000003463 573UPMBXR 00000000003107 601984XRD1 00000000003337 605322XRDER 00000000003035 609597XRD 00000000002881 610648XRD5 00000000003570 610916LIMRAD1 00000000002539 617414XRD2 00000000003178 617638DR4 00000000002842 617754XRD3 00000000002834 620429RAD 00000000002721 631727XRD 00000000003023 703698D8000 00000000001933 704384XRD 00000000003358 706733VAUD1 00000000002988 713605XRD 00000000002697 719560DR1 00000000002825 724458XRD 00000000002656 724772XRDL 000010507177AD 734467XRD 00000000003005 763780XRD 00000000002500 813571ER 00000000003044 817551OPR1 00000000002975 818907RAD1 00000000002581 847429ERT 00000000003108 847842OP1 00000000001157 850934XRD 00000000002453 859313XRD2 00000000003277 919575FCI2 00000000003448 919575LSCI 00000000002816 928779RM1 00000000003252 970247EDXRD 00000000003241 970247RADXRD 00000000003340 972981D8K6 00000000001818 972981D8K7 00000000003391 972981D8K8 00000000003700 716250CH1 00000000003437 225869XRD 00000000003424 408363VATHU1 00000000002651 602506XRD 00000000002440 615342ERXRD 00000000002996 615342RM1 00000000003289 617414LTG2 00000000003380 660827BR2 00000000003141 706494HC4 00000000003205 706571MCXR 00000000003383 706884HCL 00000000002519 715393XRD 00000000003318 724773XRDL 00000000003330 724873XRDL 00000000003388 724873XRDL2 00000000003159 732545DR1 00000000002863 781344XRD 00000000002614 805494XRD 00000000002851 847437XRDCH 00000000002884 850208XRD 00000000001544 850434XRD 00000000001161 864512XRD 00000000003272 865481FTXRD 00000000002260 734712RM1 00000000003027 903XRDFCIPRISON FMITR170990001 203573XRD FMITR170990002 205387WDC FMITR170990003 206320EHXR FMITR170990004 229312PER FMITR170990005 269657XRD FMITR170990006 309454BXRD FMITR170990007 310315SJXRD1 FMITR170990008 503315WOXRD FMITR170990009 609953XRD FMITR170990010 740592R1 FMITR170990011 757983RM1 FMITR170990012 773MTSXRDER FMITR170990013 850934XRDA FMITR170990014 931685XRD FMITR170990015 815477CLXRD FMITR170990018 920303XQI FMITR170990019 281338XDER FMITR170990020 314525ARM19A FMITR170990036 713973HDQXI FMITR170990044 ECXR1405B FMITR170990049 210297SLBXRD FMITR170990050 312996UCON FMITR170990051 312996UHUM FMITR170990052 404686CRAD1 FMITR170990053 409899RAD1 FMITR170990055 509473DR3 FMITR170990057 617726XR9Y3 FMITR170990058 650497CHXRD2 FMITR170990059 708425XRD1209 FMITR170990060 727298ER1 FMITR170990061 773967XRD3 FMITR170990062 918744JDR1 FMITR170990063 JAC1D45 FMITR170990064 JESSEBROWNADMIN FMITR170990065 404501DICXRD FMITR170990067 601883XRD FMITR170990128 216844XRD3 FMITR170990129 715833XRD FMITR170990132 415668XRD1 FMITR170990135 615264ER1 FMITR170990153 203576XRD FMITR170990154 409772RS1 FMITR170990160 615896XR2 FMITR170990161 218249XRD1 FMITR170990162 218249XRD2 FMITR170990163 847677DRAD FMITR170990167 205592XRD2 FMITR170990168 205620SXRD1 FMITR170990169 208734IMXRD FMITR170990170 210617H103 FMITR170990171 215710ER1 FMITR170990172 217876ER1 FMITR170990173 225765XRD FMITR170990174 251990XRD FMITR170990175 281929XRD1 FMITR170990176 301698XRD FMITR170990177 303695XRD FMITR170990178…
Recall started
June 4, 2014
Device type
System, X-Ray, Stationary
Root cause (FDA)
Equipment maintenance
What the company told customers
GE Healthcare sent an Urgent Medical Device Correction letter dated June 4, 2014, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. For questions in regard to this recall the consignees in the United States were advised to call 800-437-1171. For other countries contact your local GE Healthcare Service Representative. For questions regarding this recall call 800-437-1171.
Recall number
Z-1989-2014

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-1988-2014, Z-1989-2014. The agency record governs; check it and the company before acting.