Menu
FDA · Medical device recall

EndoGastric Solutions EsophyX2 Device with Serosa Fastener and Accessories…

Aug. 28, 2013 Class II

Date
Aug. 28, 2013
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Endogastric Solutions Inc
Product
EndoGastric Solutions EsophyX2 Device with Serosa Fastener and Accessories. Models R2001 and R2002. Indicated for the treatment of symptomatic chronic gastroesophageal reflux disease.
Reason
Endogastric Solutions, Inc. has received a limited number of reports relating to the loss of tissue mold control when operating the R2001 or R2002 EsophyX2 device. In one case, surgical intervention was necessary to remove the device.
Distribution pattern
Distributed nationwide and Italy.
Quantity
5192 units in the US and 41 units outside the US
Codes and lots
UN-EXPIRED devices with Expiration dates (8/31/2013, 11/30/2013, 12/31/2013, and 1/31/2014) and following LOT numbers: 401132, 401147, 401152, 401156, 401159, 401167, 401207, 401214, 401219, 401220, 401225, 401227, 401230, 401233, 401236, 401246, 401248, 401251, 401257, 401258, 401259, 401267, 401269, 401271, 401274, and 401276. EXPIRED devices with Expired dates (1/31/2013, 2/28/2013, 3/31/2013 4/30/2013, 5/31/2013, 6/30/2013, and 7/31/2013) and following LOT numbers: 400923, 400927, 400934, 400937, 400940, 400943, 400949, 400961, 400963, 400973, 400976, 400979, 400980, 400981, 400984, 400986, 400992, 400994, 400997, 400998, 401004, 401005, 401009, 401012, 401016, 401017, 401024, 401026, 401029, 401056, 401058, 401059, 401060, 401061, 401062, 401063, 401065, 401072, 401074, 401076, 401079, 401084, 401085, 401092, 401094, 401107, 401114, 401115, 401121, 401129, and 401130.
Recall started
June 6, 2013
Device type
Endoscopic Suture/Plication System, Gastroesophageal Reflux Disease (Gerd)
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Endogastric sent the Safety Alert: EndoGastric Solutions (EGS), EsophyX2 Device letter, dated June 5, 2013, to their consignees. Endogastric sent the second letter URGENT: MEDICAL DEVICE RECALL EsophyX2, dated July 26, 2013. This letter advised customers that the firm is voluntarily recalling EsophyX2 Devices with SerosaFuse Implantable Fastener and Accessories (Models R2001 and R2002) manufactured before February 2012. Customers who have UN-EXPIRED devices in their inventory are advised to discontinue use and fill out the Medical Device Recall Return Response form and return it to Endogastric. The Customer Service will contact customers with instructions on how to return the product to the company. Customers who may have EXPIRED devices are advised to with their central supply departments to ensure all identified products have been removed from inventory and destroyed. They should fill out the Medical Device Recall Return Response form with the lot numbers and quantity destroyed and return the completed form to the firm. Customers can call the Customer Service and Support at 425-307-9269, Monday through Friday, 8:00AM to 5:00 PM, Pacific Time for questions. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Recall number
Z-1982-2013