FDA · Medical device recall
Precision 500D X-ray systems with under-table collimator part number 5234960…
- Date posted by the FDA
- May 6, 2011
- Status
- Terminated
- Filed by
- GE Healthcare, LLC
- Product
- Precision 500D X-ray systems with under-table collimator part number 5234960. The Precision 500D is designed to perform radiographic and fluoroscopic x-ray examinations.
- Reason
- Use of high force when moving the image carriage/IDD into forward/park position can cause changes to the mechanical alignment of the collimator blades and allow the blades to open wider than specified, consequently allowing radiation outside the permitted field without warning to the user. This issue manifests with symptoms wherein the user cannot fully close down the lateral blades.
- Distribution pattern
- Nationwide Distribution
- Quantity
- 440 US
- Codes and lots
- Part number 5234960.
- Recall started
- June 10, 2010
- Device type
- System, X-Ray, Fluoroscopic, Image-Intensified
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- GE Healthcare sent an Importan Electronic Product Radiation Warning letter to customers. The letter discussed the product defect description and related hazards, affected product details, instructions, and product correction. The firm will send a GE Field Service Representative to visit each affected customer site to inspect and correct all potentially affected units in the field following the instructions in field modification instructions No: 10846. If you have any questions please call 800-437-1171.
- Recall number
- Z-1971-2011
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-1971-2011. The agency record governs; check it and the company before acting.