Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308…

June 18, 2025 Class II

Date
June 18, 2025
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Ortho-Clinical Diagnostics, Inc.
Product
Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software Version: not applicable Product Description: One sales unit (kit) contains materials for 48 tests. One kit contains 48 tubes of 1.45 mL Process Buffer and 48 Reaction Tubes. Component: No, not a component
Reason
It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.
Distribution pattern
Please see the attached list "Consignee List_CL2025-120." Note: This file contains a specific tab, titled Shipping Summary, which contains the US and ex-US shipping summary. One (1) Distributor was affected. The other four(4) Distributors were sent the notification for information only. QuidelOrtho sent the notification to their consignees because QuidelOrtho drop-shipped this product to their 31 consignees (Included in the Medical Facility total below).
Quantity
952 units
Codes and lots
Lot Code: Catalog number: M308 UDI-DI: 30014613336167; Lots: 245251 - Exp. 2025-Oct-25 246941 - Exp. 2025-Dec-03 250163 - Exp. 2026-Feb-06 259686 - Exp. 2026-Sep-03 261072 - Exp. 2026-Oct-02 262113 - Exp. 2026-Oct-23 263331 - Exp. 2026-Nov-22 264065 - Exp. 2026-Dec-11 265007 - Exp. 2027-Jan-03 (corrected exp date from 2026-Dec-31) 266928 - Exp. 2026-Dec-31 Additional Lots: 265007-1 - Exp. 2026-Dec-31 268479 and above - Various Expiration dates *Until further notice, all future lots, both supplied by QuidelOrtho or supplied by its distributors, will be affected.
Recall started
March 24, 2025
Device type
Bordetella Pertussis Dna Assay System
Root cause (FDA)
Under Investigation by firm
What the company told customers
On May 6, 2025, URGENT PRODUCT CORRECTION NOTIFICATION letters to customers. REQUIRED ACTION " Do not report Bordetella parapertussis (BPP) positive results when testing with the affected lot numbers. " Use the Solana Bordetella Complete Assay only for reporting Bordetella pertussis (BP) results and negative Bordetella parapertussis (BPP) results only when using the affected lots, as they are not affected. " Review the content of this communication with your Medical Director and retain this letter for your laboratory. " Complete the enclosed Confirmation of Receipt form no later than May 14, 2025. Contact Information We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact our Global Services Organization at 1-800-874-1517 or technicalsupport@quidelortho.com.
Recall number
Z-1948-2025

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-1948-2025. The agency record governs; check it and the company before acting.