FDA · Medical device recall
Disposable Drill Kit which includes two drill bits…
- Date
- June 6, 2018
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Ad-Tech Medical Instrument Corporation
- Product
- Disposable Drill Kit which includes two drill bits, two drill stops, and two adjustment wrenches, sterile. The drill bits are 30cm long and made of stainless steel. The drill kits are supplied in a sterile state and are single-use only. Used to drill cranial holes using a stereotactic frame.
- Reason
- There is a possibility that DDK2-2.4-30X Disposable Drill Kits, Lot Number 111664 208140649 contained 2.8mm drill bits from DDK2-2.8-30X Disposable Drill Kits, Lot Number 111745 208140649 and vice versa.
- Distribution pattern
- Worldwide Distribution: US (nationwide) to states of: CA, DC, FL, MI, MA, NC, NY, WA, WI; and country of: Canada. Distributed internationally to the United Kingdom.
- Quantity
- 30 devices
- Codes and lots
- Catalog No. DDK2-2.4-30X, Lot No. 0111664 208140649, UDI No. 01008418231011741719080110011174 5208140649; Catalog No. DDK2-2.8-30X, Lot No. 0111745 208140649, UDI No. 01008418231011671719080110011166 4208140649
- Recall started
- Dec. 20, 2017
- Device type
- Orthopedic Manual Surgical Instrument
- Root cause (FDA)
- Packaging process control
- What the company told customers
- The firm, Ad-Tech Medical Instrument Corporation, sent a "MEDICAL DEVICE RECALL" letter on approximately 12/20/2017 to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to examine inventory and quarantine any affected product on hand; notify customers of the recall if affected product was further distributed; complete and return the Acknowledgement and Receipt Form via Fax to: 262-634-5668, Attention Customer Support; Mail to: AD-TECH MEDICAL INSTRUMENT CORPORATION, 1901 WILLIAM STREET, RACINE, WI 53404 or Email to: customersupport@adtechmedical.com, and contact an Ad-Tech Customer Support Specialist for a Return Material Authorization (RMA) number. If you have any further questions, contact Regulatory Team Representative and CAPA/Complaints Manager, Monday through Friday, 7:00 AM to 5:00 PM, Central Time via FAX: 262-634-5668, Toll Free: 1-800-776-1555 or Email: customersupport@adtechmedical.com.
- Recall number
- Z-1938-2018
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-1938-2018. The agency record governs; check it and the company before acting.