FDA · Medical device recall
Chiba Needle Marked Model: CNM-2308…
- Date posted by the FDA
- Sept. 20, 2009
- Status
- Terminated
- Filed by
- Remington Medical Inc.
- Product
- Chiba Needle Marked Model: CNM-2308, Size: 23 GA x 20 CM, Remington Medical, Inc., Alpharetta, GA 30005. General purpose aspiration Cytology Biopsy Needle
- Reason
- Two unsealed pouches were identified after product release, compromising product sterility.
- Distribution pattern
- US - CA and WI.
- Quantity
- 50 units
- Codes and lots
- Lot number 091386.
- Recall started
- June 19, 2009
- Root cause (FDA)
- Process control
- What the company told customers
- Initial notification to consignees was issued via telephone on June 19, 2009 and by certified mail on June 30, 2009. They were instructed to examine their inventory and quarantine product subject to the recall. If the product was further distributed, they were advised to identify their customers and notify them at once of the recall via the recall letter. A Return Authorization number was provided. Forty of 50 units in distribution were returned from the two customers.
- Recall number
- Z-1935-2009