Case Cardiac Assessment System for Exercise Testing and CardioSoft Diagnostic System Exercise Stress Testing…

June 30, 2021 Class II

Date
June 30, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Product
Case Cardiac Assessment System for Exercise Testing and CardioSoft Diagnostic System Exercise Stress Testing ECG application
Reason
If a certain sequence of events occur, the pdf test report that belongs to one patient will appear in a different patient's record when viewed in an Electronic Medical Record (EMR) or a similar system.
Distribution pattern
Worldwide distribution - US Nationwide. Government/military distribution was also made, and the countries of Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Brazil, Brunei Darussalam, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czech Republic, Czechia, Denmark, Estonia, Fiji, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Malta, Mexico, Moldova, Myanmar, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Viet Nam.
Quantity
11,295 devices (5,637 US; 5,658 OUS)
Codes and lots
CASE and CardioSoft - v6.6x, v6.7x and v7.0
Recall started
April 19, 2021
Device type
Computer, Diagnostic, Programmable
Root cause (FDA)
Software design
What the company told customers
The recalling firm issued letters dated 4/19/2021 beginning 4/19/2021. For systems that are on Windows 8.1 and above, the issue will be corrected through a software patch. For the units running on Microsoft Windows Server 2008-R2, Windows XP, and Windows 7, the letter instructs that the devices are not to be networked on these systems and that a software patch cannot be made available to these systems since the operating systems are not supported by Microsoft. A second cover letter and customer notification were issued dated 5/10/2021 updating the second safety instruction regarding the units running on Microsoft Windows Server 2008-R2, Windows XP, and Windows 7 saying that these systems must be rebooted before manually exporting tests.
Recall number
Z-1927-2021

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-1927-2021. The agency record governs; check it and the company before acting.