FDA · Medical device recall
Zoll Auto Pulse Resuscitation System…
- Date posted by the FDA
- April 7, 2011
- Status
- Terminated
- Filed by
- Zoll Circulation, Inc.
- Product
- Zoll Auto Pulse Resuscitation System; Zoll Circulation Inc., Sunnyvale, CA 94085 Cardiac Compressor used as an adjunct to manual CPR. Use of AutoPulse is intended to reduce the impact of rescuer fatigue and enable rescuer to address other patient needs.
- Reason
- Complaints of batteries discharging prematurely and that during use the Auto Pulse Units would cease operating (providing compressions).
- Distribution pattern
- Nationwide Distribution -- including, VA, NY, OH, IN, MO, UT, FL, NC, ME, WA, SD, NM, PA, MS, ID, TX, GA, CA, CO, IL, AR, WV, TN, OR, and KS.
- Quantity
- not provided
- Codes and lots
- Model 100; AutoPulse product number 8700-0700-xx AutoPulse 1.5G; AutoPulse Battery Part Number 8799-0702-xx
- Recall started
- Jan. 28, 2008
- Device type
- Compressor, Cardiac, External
- Root cause (FDA)
- Other
- What the company told customers
- The firm, Zoll Circulation Inc., initiated a Corrective and Preventive Action Report on January 28, 2009. According to the firm, the customers were contacted by telephone calls and were provided with a Battery Maintenance program. For further questions you may call (408) 541-2140 ext.341.
- Recall number
- Z-1904-2011