FDA · Medical device recall

Revolution EVO…

July 7, 2021 Class II

Date
July 7, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
GE Healthcare, LLC
Product
Revolution EVO, Optima CT660, Optima CT680 CT Systems.
Reason
Improperly loaded software options may result in additional X -ray radiation exposure to the patient.
Distribution pattern
3,665 (US = 956; OUS = 2,709)
Quantity
3665
Codes and lots
Optima CT660 12HW28.8 14HW30.3 15HW25.2 13HW38.14 15HW13.7 Revolution EVO 14HW17.4 15HW13.7 Optima CT680 13HW38.14 15BW21.2
Recall started
June 14, 2021
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare will send an Important Electronic Product Radiation Warning letter to all affected systems informing its users of the potential for the issue to occur and instruct the users as to what actions to take to prevent this issue.
Recall number
Z-1900-2021

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-1900-2021. The agency record governs; check it and the company before acting.