FDA · Medical device recall
Atec Insignia Anterior Cervical Plate System… and 28 more products
- Date
- June 16, 2021
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Alphatec Spine, Inc.
- Reason
- There is a potential for the screw blocking mechanism on the anterior cervical plate system to disassociate intraoperatively/postoperatively or unlock postoperatively.
- Distribution pattern
- US Nationwide distribution in the states of FL, CA, IL, OK, TX, CT, IN, NC, WA, HI, ID, MA, SC, NM, and NY.
29 products in this recall
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0116, Insignia, ACP, 1-Level, 16 mm, Rx Only, Non-Sterile, UDI: (01)00190376268422
- Quantity
- 8 units
- Codes and lots
- 8609301R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1799-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0117, Insignia, ACP, 1-Level, 17 mm, Rx Only, Non-Sterile, UDI: (01)00190376268439
- Quantity
- 6 units
- Codes and lots
- 8663405R and 8800402
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1800-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0118, Insignia, ACP, 1-Level, 18 mm, Rx Only, Non-Sterile, UDI: (01)00190376268446
- Quantity
- 7 units
- Codes and lots
- 8609303R and 8800403
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1801-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0119, Insignia, ACP, 1-Level, 19 mm, Rx Only, Non-Sterile, UDI: (01)00190376268453
- Quantity
- 13 units
- Codes and lots
- 8609304R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1802-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0120, Insignia, ACP, 1-Level, 20 mm, Rx Only, Non-Sterile, UDI: (01)00190376268460
- Quantity
- 10 units
- Codes and lots
- 8609305R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1803-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0121, Insignia, ACP, 1-Level, 21 mm, Rx Only, Non-Sterile, UDI: (01)00190376268477
- Quantity
- 7 units
- Codes and lots
- 8609306R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1804-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0122, Insignia, ACP, 1-Level, 22 mm, Rx Only, Non-Sterile, UDI: (01)00190376268484
- Quantity
- 8 units
- Codes and lots
- 8609307R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1805-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0124, Insignia, ACP, 1-Level, 24 mm, Rx Only, Non-Sterile, UDI: (01)00190376268743
- Quantity
- 8 units
- Codes and lots
- 8609309R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1806-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0232 Insignia, ACP, 2-Level, 32 mm, Rx Only, Non-Sterile, UDI: (01)00190376268361
- Quantity
- 3 units
- Codes and lots
- 8609312R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1807-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0234 Insignia, ACP, 2-Level, 34 mm, Rx Only, Non-Sterile, UDI: (01)00190376268378
- Quantity
- 6 units
- Codes and lots
- 8609313R and 8800410
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1808-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0236 Insignia, ACP, 2-Level, 36 mm, Rx Only, Non-Sterile, UDI: (01)00190376268385
- Quantity
- 18 units
- Codes and lots
- 8609314R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1809-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0238 Insignia, ACP, 2-Level, 38 mm, Rx Only, Non-Sterile, UDI: (01)00190376268392
- Quantity
- 19 units
- Codes and lots
- 8609315R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1810-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0240 Insignia, ACP, 2-Level, 40 mm, Rx Only, Non-Sterile, UDI: (01)00190376268408
- Quantity
- 26 units
- Codes and lots
- 8663401R, 8609316R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1811-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0242 Insignia, ACP, 2-Level, 42 mm, Rx Only, Non-Sterile, UDI: (01)00190376268415
- Quantity
- 14 units
- Codes and lots
- 8663402R and 8609317R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1812-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0244 Insignia, ACP, 2-Level, 44 mm, Rx Only, Non-Sterile, UDI: (01)00190376268507
- Quantity
- 3 units
- Codes and lots
- 8663403R & 8609318R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1813-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0248 Insignia, ACP, 2-Level, 48 mm, Rx Only, Non-Sterile, UDI: (01)00190376268521
- Quantity
- 3 units
- Codes and lots
- 8609319R and 8663404R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1814-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0250 Insignia, ACP, 2-Level, 50 mm, Rx Only, Non-Sterile, UDI: (01)00190376268538
- Quantity
- 1 unit
- Codes and lots
- 8663405R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1815-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0350 Insignia, ACP, 3-Level, 50 mm, Rx Only, Non-Sterile, UDI: (01)00190376268569
- Quantity
- N/A
- Codes and lots
- 8609322R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1816-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0352 Insignia, ACP, 3-Level, 52 mm, Rx Only, Non-Sterile, UDI: (01)00190376268576
- Quantity
- 4 units
- Codes and lots
- 8609323R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1817-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0354 Insignia, ACP, 3-Level, 54 mm, Rx Only, Non-Sterile, UDI: (01)00190376268583
- Quantity
- 7 units
- Codes and lots
- 8609324R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1818-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0356 Insignia, ACP, 3-Level, 56 mm, Rx Only, Non-Sterile, UDI: (01)00190376268590
- Quantity
- 5 units
- Codes and lots
- 8609325R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1819-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0358 Insignia, ACP, 3-Level, 58 mm, Rx Only, Non-Sterile, UDI: (01)00190376268606
- Quantity
- 3 units
- Codes and lots
- 8609326R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1820-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0360 Insignia, ACP, 3-Level, 60 mm, Rx Only, Non-Sterile, UDI: (01)00190376268613
- Quantity
- 1 units
- Codes and lots
- 8609327R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1821-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0362 Insignia, ACP, 3-Level, 62 mm, Rx Only, Non-Sterile, UDI: (01)00190376268620
- Quantity
- 7 units
- Codes and lots
- 8609328R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1822-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0364 Insignia, ACP, 3-Level, 64 mm, Rx Only, Non-Sterile, UDI: (01)00190376268637
- Quantity
- 1 units
- Codes and lots
- 8609329R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1823-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0468 Insignia, ACP, 4-Level, 68 mm, Rx Only, Non-Sterile, UDI: (01)00190376268675
- Quantity
- 2 units
- Codes and lots
- 8609331R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1824-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0474 Insignia, ACP, 4-Level, 74 mm, Rx Only, Non-Sterile, UDI: (01)00190376268699
- Quantity
- 5 units
- Codes and lots
- 8609333R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1825-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0478 Insignia, ACP, 4-Level, 78 mm, Rx Only, Non-Sterile, UDI: (01)00190376268705
- Quantity
- 1 unit
- Codes and lots
- 8609334R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1826-2021
Atec Insignia Anterior Cervical Plate System…
- Product
- Atec Insignia Anterior Cervical Plate System, REF 136-0482 Insignia, ACP, 4-Level, 82 mm, Rx Only, Non-Sterile, UDI: (01)00190376268712
- Quantity
- 1 unit
- Codes and lots
- 8609335R
- Recall started
- April 26, 2021
- Device type
- Appliance, Fixation, Spinal Intervertebral Body
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Between the dates of 4/27-28/2021, Alphatec Spine, Inc., distributed a "Urgent-Medical Device Recall Notification" . The notification was distributed via FedEx and E-Mail. In addition to informing consignees about the recall, Alphatec Spine Inc. asked customers to take the following actions: 1. Please promptly review your inventory to determine if any of the affected product listed above is within your possession. 2. Please abstain from sale and/or use and contact Alphatec Spine Customer Service for instructions on how to carry out a product return. All shipping instructions will be provided, including arrangements for product replacement. 3. Please fill out the last page of this letter to confirm that you have read this notification and have taken all necessary actions as described in this notification. 4. Please return page 4 (signed) of this letter using one of the methods below: Mail to: Alphatec Spine, Inc., 1950 Camino Vida Roble, Carlsbad, CA 92008 ATTN: Hylene Valdez Email to: QMScompliance@atecspine.com 5. Please share this notice with all those who need to be aware within your organization or any organization where the affected product has been transferred. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to Alphatec at 1-800-922-1356 or 1-760-431-9286. The FDAs MedWatch Adverse Event reporting program can also be used, via email, on-line or fax.
- Recall number
- Z-1827-2021
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement reports, recall numbers Z-1799-2021, Z-1800-2021, Z-1801-2021, Z-1802-2021, Z-1803-2021, Z-1804-2021, Z-1805-2021, Z-1806-2021 and others. The agency record governs; check it and the company before acting.