FDA · Medical device recall

Discovery PET/CT 610 Product Usage: The systems are intended for head… and 5 more products

June 26, 2019 Class II

Date
June 26, 2019
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
GE Healthcare, LLC
Reason
GE Healthcare PET/CT Table Stop Without Stopping X-Ray Exposure
Distribution pattern
Nationwide Distribution

6 products in this recall

Discovery PET/CT 610 Product Usage: The systems are intended for head and whole body attenuation corrected…

Product
Discovery PET/CT 610 Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.
Quantity
1113 in total
Codes and lots
CT 610
Recall started
Feb. 12, 2019
Device type
System, Tomography, Computed, Emission
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare sent a notification letter to affected customer (Hospital Administrators/Risk Managers. Managers of Radiology/Cardiology, Radiologist/Cardiologist). The letter identified the affected product, problem and actions to be taken. Customers were informed GE Healthcare will without charge, remedy the defect or bring the product into compliance. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1745-2019

Discovery PET/CT 710 Product Usage: The systems are intended for head and whole body attenuation corrected…

Product
Discovery PET/CT 710 Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.
Quantity
1113 in total
Codes and lots
CT 710
Recall started
Feb. 12, 2019
Device type
System, Tomography, Computed, Emission
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare sent a notification letter to affected customer (Hospital Administrators/Risk Managers. Managers of Radiology/Cardiology, Radiologist/Cardiologist). The letter identified the affected product, problem and actions to be taken. Customers were informed GE Healthcare will without charge, remedy the defect or bring the product into compliance. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1746-2019

Optima PET/CT 560…

Product
Optima PET/CT 560, 560FX Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.
Quantity
1113 in total
Codes and lots
CT 560, 560FX
Recall started
Feb. 12, 2019
Device type
System, Tomography, Computed, Emission
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare sent a notification letter to affected customer (Hospital Administrators/Risk Managers. Managers of Radiology/Cardiology, Radiologist/Cardiologist). The letter identified the affected product, problem and actions to be taken. Customers were informed GE Healthcare will without charge, remedy the defect or bring the product into compliance. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1747-2019

Discovery IQ Product Usage: The systems are intended for head and whole body attenuation corrected Positron…

Product
Discovery IQ Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.
Quantity
1113 in total
Codes and lots
Discovery IQ
Recall started
Feb. 12, 2019
Device type
System, Tomography, Computed, Emission
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare sent a notification letter to affected customer (Hospital Administrators/Risk Managers. Managers of Radiology/Cardiology, Radiologist/Cardiologist). The letter identified the affected product, problem and actions to be taken. Customers were informed GE Healthcare will without charge, remedy the defect or bring the product into compliance. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1748-2019

Discovery Ml Product Usage: The systems are intended for head and whole body attenuation corrected Positron…

Product
Discovery Ml Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.
Quantity
1113 in total
Codes and lots
Discovery Ml
Recall started
Feb. 12, 2019
Device type
System, Tomography, Computed, Emission
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare sent a notification letter to affected customer (Hospital Administrators/Risk Managers. Managers of Radiology/Cardiology, Radiologist/Cardiologist). The letter identified the affected product, problem and actions to be taken. Customers were informed GE Healthcare will without charge, remedy the defect or bring the product into compliance. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1749-2019

Discovery Ml DR Product Usage: The systems are intended for head and whole body attenuation corrected…

Product
Discovery Ml DR Product Usage: The systems are intended for head and whole body attenuation corrected Positron Emission Tomography (PET) imaging and localization of emission activity in patient anatomy by means of integrated PET and CT images.
Quantity
1113 in total
Codes and lots
Discovery Ml DR
Recall started
Feb. 12, 2019
Device type
System, Tomography, Computed, Emission
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare sent a notification letter to affected customer (Hospital Administrators/Risk Managers. Managers of Radiology/Cardiology, Radiologist/Cardiologist). The letter identified the affected product, problem and actions to be taken. Customers were informed GE Healthcare will without charge, remedy the defect or bring the product into compliance. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1750-2019

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-1745-2019, Z-1746-2019, Z-1747-2019, Z-1748-2019, Z-1749-2019, Z-1750-2019. The agency record governs; check it and the company before acting.