Lunar iDXA bone densitometer and 3 more products

May 6, 2020 Class II

Date
May 6, 2020
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
GE Healthcare, LLC
Reason
An issue with the Lunar iDXA control panel may cause the start button to not release and potentially contribute to an accidental radiation occurrence.
Distribution pattern
US Nationwide Distribution

4 products in this recall

Lunar iDXA bone densitometer

Product
Lunar iDXA bone densitometer
Quantity
123 in total
Codes and lots
all affected systems
Recall started
Dec. 3, 2019
Device type
Densitometer, Bone
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare notified customers via an Electronic Product Radiation Warning Letter dated March 13, 2020. The letter provided the following instructions. 1) Discontinue use of the start button on the control panel of the Lunar iDXA system to start a scan until a GE Representative can correct your system. 2) Disable the control panel start button in the enCORE software by following the below steps: " In enCORE, select menu item Tools> User Options. " Select the measure tab. " Uncheck the option "Allow Scanner Start Button to initiate a Measurement " " Press OK. 3) Start the patient scan from the measure screen of the enCORE software. GE Healthcare will correct all affected Lunar iDXA systems by inspecting and replacing, if necessary, the iDXA control panel at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1709-2020

Lunar iDXA Advance (iDXA with enCORE v16) bone densitometer

Product
Lunar iDXA Advance (iDXA with enCORE v16) bone densitometer
Quantity
123 in total
Codes and lots
all affected systems
Recall started
Dec. 3, 2019
Device type
Densitometer, Bone
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare notified customers via an Electronic Product Radiation Warning Letter dated March 13, 2020. The letter provided the following instructions. 1) Discontinue use of the start button on the control panel of the Lunar iDXA system to start a scan until a GE Representative can correct your system. 2) Disable the control panel start button in the enCORE software by following the below steps: " In enCORE, select menu item Tools> User Options. " Select the measure tab. " Uncheck the option "Allow Scanner Start Button to initiate a Measurement " " Press OK. 3) Start the patient scan from the measure screen of the enCORE software. GE Healthcare will correct all affected Lunar iDXA systems by inspecting and replacing, if necessary, the iDXA control panel at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1710-2020

Lunar iDXA Pro/Lunar iDXA Advance (iDXA with enCORE v17) bone densitometer

Product
Lunar iDXA Pro/Lunar iDXA Advance (iDXA with enCORE v17) bone densitometer
Quantity
123 in total
Codes and lots
all affected systems
Recall started
Dec. 3, 2019
Device type
Densitometer, Bone
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare notified customers via an Electronic Product Radiation Warning Letter dated March 13, 2020. The letter provided the following instructions. 1) Discontinue use of the start button on the control panel of the Lunar iDXA system to start a scan until a GE Representative can correct your system. 2) Disable the control panel start button in the enCORE software by following the below steps: " In enCORE, select menu item Tools> User Options. " Select the measure tab. " Uncheck the option "Allow Scanner Start Button to initiate a Measurement " " Press OK. 3) Start the patient scan from the measure screen of the enCORE software. GE Healthcare will correct all affected Lunar iDXA systems by inspecting and replacing, if necessary, the iDXA control panel at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1711-2020

Lunar iDXA Pro/Lunar iDXA Advance (iDXA with enCORE v18) bone densitometer

Product
Lunar iDXA Pro/Lunar iDXA Advance (iDXA with enCORE v18) bone densitometer
Quantity
123 in total
Codes and lots
all affected systems
Recall started
Dec. 3, 2019
Device type
Densitometer, Bone
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare notified customers via an Electronic Product Radiation Warning Letter dated March 13, 2020. The letter provided the following instructions. 1) Discontinue use of the start button on the control panel of the Lunar iDXA system to start a scan until a GE Representative can correct your system. 2) Disable the control panel start button in the enCORE software by following the below steps: " In enCORE, select menu item Tools> User Options. " Select the measure tab. " Uncheck the option "Allow Scanner Start Button to initiate a Measurement " " Press OK. 3) Start the patient scan from the measure screen of the enCORE software. GE Healthcare will correct all affected Lunar iDXA systems by inspecting and replacing, if necessary, the iDXA control panel at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1712-2020

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-1709-2020, Z-1710-2020, Z-1711-2020, Z-1712-2020. The agency record governs; check it and the company before acting.