FDA · Medical device recall
Integra 3…
- Date
- May 4, 2016
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Research Instruments Ltd - Site B
- Product
- Integra 3; Model Numbers: 6-54-100, 6-54-110, 6-54-150, 6-54-160 Product Usage: The Rl TDU500 Micromanipulator for Assisted Reproduction is used to accurately position microtools under a microscope for the techniques of Assisted Reproduction. The Rl Integra is a configuration comprising two TDU500 micromanipulators, two air syringes and temperature control system within a single housing.
- Reason
- The units have been fitted with a potentially faulty power supply cable.
- Distribution pattern
- Worldwide Distribution - US Nationwide in the states of: MA, NY and countries of : Algeria, Australia, Austria, Bahrain, Belgium, British Virgin Islands, Bulgaria, Canada, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Korea, Kuwait, Libya, Lithuania, Macedonia, Malaysia, Malta, Mexico, Nigeria, Norway, Pakistan, Peru, Poland, Romania, Serbia, Singapore, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Thailand, Tunisia, Turkey, UAE, Ukraine, United Kingdom, and Vietnam,
- Quantity
- 380 units
- Codes and lots
- Serial numbers: 2-381
- Recall started
- March 11, 2016
- Device type
- Micromanipulators And Microinjectors, Assisted Reproduction
- Root cause (FDA)
- Device Design
- What the company told customers
- Research Instruments sent a recall notification dated 03/11/2016, via UPS with confirmed delivery receipt, to all affected customers. The letter identified the affected product, problem and the action needed to be taken by the customer. The letter asked customers to contact Research Instruments First Line Support if they had any questions.
- Recall number
- Z-1584-2016
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-1584-2016. The agency record governs; check it and the company before acting.