NM/CT 850 Nuclear Medicine / CT Scanners and 2 more products

May 19, 2021 Class II

Date
May 19, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
GE Healthcare, LLC
Reason
GE Healthcare has become aware of a potential issue on 800 family NM/CT Scanners that could cause a shift in CT radiation exposure range of up to 5cm compared to the intended radiation exposure range of the planned scan under specific workflows. The issue occurs only on a hybrid whole-body continuous F3 protocol with Zoom < 1 where the scan range is set on the Smart Console. In some cases, this may also necessitate a re-scan of the patient, which would expose the patient to additional X-ray radiation.
Distribution pattern
Worldwide distribution - US Nationwide

3 products in this recall

NM/CT 850 Nuclear Medicine / CT Scanners

Product
NM/CT 850 Nuclear Medicine / CT Scanners
Quantity
242 in total (17 units US)
Codes and lots
GTIN*00840682140775
Recall started
Feb. 7, 2021
Device type
System, Tomography, Computed, Emission
Root cause (FDA)
Device Design
What the company told customers
GE Healthcare will send a customer notification letter describing the issue and providing safety instructions for immediate mitigation of the issue. Customers were instructed, they may continue to use the system and follow workaround instructions (zoom=1 for specific WB protocol) until the correction is provided. GE Healthcare corrective action is the development and deployment of a software update on all affected smart console IB systems at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1545-2021

NM/CT 860 Nuclear Medicine / CT Scanners

Product
NM/CT 860 Nuclear Medicine / CT Scanners
Quantity
242 in total (17 units US)
Codes and lots
GTIN*00840682140751
Recall started
Feb. 7, 2021
Device type
System, Tomography, Computed, Emission
Root cause (FDA)
Device Design
What the company told customers
GE Healthcare will send a customer notification letter describing the issue and providing safety instructions for immediate mitigation of the issue. Customers were instructed, they may continue to use the system and follow workaround instructions (zoom=1 for specific WB protocol) until the correction is provided. GE Healthcare corrective action is the development and deployment of a software update on all affected smart console IB systems at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1546-2021

NM/CT 870 DR Nuclear Medicine / CT Scanners

Product
NM/CT 870 DR Nuclear Medicine / CT Scanners
Quantity
242 in total (17 units US)
Codes and lots
GTIN*00840682140836
Recall started
Feb. 7, 2021
Device type
System, Tomography, Computed, Emission
Root cause (FDA)
Device Design
What the company told customers
GE Healthcare will send a customer notification letter describing the issue and providing safety instructions for immediate mitigation of the issue. Customers were instructed, they may continue to use the system and follow workaround instructions (zoom=1 for specific WB protocol) until the correction is provided. GE Healthcare corrective action is the development and deployment of a software update on all affected smart console IB systems at no cost to the customer. For questions contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-1547-2021

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-1545-2021, Z-1546-2021, Z-1547-2021. The agency record governs; check it and the company before acting.