FDA · Medical device recall
VITROS¿ Black Reference Slides CATALOG #: J02316…
- Date
- May 5, 2021
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Ortho-Clinical Diagnostics, INc.
- Product
- VITROS¿ Black Reference Slides CATALOG #: J02316 - Product Usage: The Black Reference Slide is a part/ tool used in the reflectometer correction factor adjustment procedure.
- Reason
- Delayed results. Ortho confirmed that two lots of VITROS Black Reference Slides are incorrectly labelled as White Reference Slides. If the customer fails to notice both the product name or the unexpected color of the slides and tries to use Black Reference Slides as White Reference Slides, the Correction Factor test will fail - and the analyzer will not be able to produce results for any assays other than sodium, potassium, and chloride. If the customer doesnt have any White Reference Slides in reserve, they cant run the Correction Factor test correctly and get the analyzer running again until they get more White Reference Slides delivered. This could lead to a delay in results for all Vitros assays except for sodium, potassium, and chloride.
- Distribution pattern
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, CT, FL GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, ND, NE, NH, NJ, NY, OH, OK, PA, SC, TN, TX, VA, WA, WI, WV, WY and the countries of Australia, Bermuda, Brazil, Canada, China, Colombia, India, Japan, Mexico, Singapore, United Kingdom, France, Germany, Italy, Spain, Portugal, Poland, Russia, Denmark, Norway, Sweden, Belgium, The Netherlands.
- Quantity
- 865 units of 25 slides each
- Codes and lots
- Lots: 9252-0058-3361, 9252-0058-3362
- Recall started
- March 26, 2021
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On 3/25/2021, Ortho Clinical Diagnostics issued an Urgent Medical Device Correction notice via letter notifying customers two lots of the VITROS Black Reference Slides have the incorrect part number on the box.
- Recall number
- Z-1489-2021
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-1489-2021. The agency record governs; check it and the company before acting.