FDA · Medical device recall
ANGIOGUARD RX / XP Emboli Capture Guidewire System… and 13 more products
- Date
- May 17, 2023
- Risk class
- Class I (FDA's definition, 21 CFR 7.3(m): a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death)
- Status
- Ongoing
- Filed by
- Cordis US Corp
- Reason
- There is a potential for separation of the ANGIOGUARD RX / XP delivery system and capture sheath which include but are not limited to situations of an intra-procedural delay, unplanned percutaneous or surgical intervention, or stroke; while a replacement device is prepared.
- Distribution pattern
- Worldwide distribution.
14 products in this recall
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 401814RMC
- Quantity
- 283 units
- Codes and lots
- UDI/DI 20705032053508; Lot Numbers: 35262517, 35263334, 35264204, 35264222, 35265345.
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1483-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814RMC
- Quantity
- 684 units
- Codes and lots
- UDI/DI 20705032054802; Lot Numbers: 35264208, 35264212, 35265329, 35265654, 35265649, 35265667.
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1484-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 601814RMC
- Quantity
- 471 units
- Codes and lots
- UDI/DI 20705032056745; Lot Numbers: 35263328, 35264207, 35264213, 35264216, 35265393, 35265648, 35265659
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1485-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 701814RMC
- Quantity
- 54 units
- Codes and lots
- UDI/DI 20705032056851; Lot Numbers: 35265661
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1486-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 801814RMC
- Quantity
- 146 units
- Codes and lots
- UDI/DI 20705032056943; Lot Numbers: 35265335, 35264202
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1487-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 603014MC
- Quantity
- 43 units
- Codes and lots
- UDI/DI 20705032056790; Lot Numbers: 35265492
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1488-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814REC
- Quantity
- 40 units
- Codes and lots
- UDI/DI 20705032054789; Lot Numbers: 35264223
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1489-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 401814RM
- Quantity
- 56 units
- Codes and lots
- UDI/DI 20705032053492; Lot Numbers: 35265339, 35265670
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1490-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814RE
- Quantity
- 1087 units
- Codes and lots
- UDI/DI 20705032054772; Lot Numbers: 35264217, 35264226, 35265330, 35265344, 35265381, 35265641, 35265652, 35265655, 35264205, 35264210, 35264214, 35265340, 35265341, 35265394, 35265643
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1491-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 601814RE
- Quantity
- 1058 units
- Codes and lots
- UDI/DI 20705032056714; Lot Numbers: 35264218, 35264224, 35264806, 35265342, 35265343, 35265382, 35265383, 35265646, 35265656, 35265658, 35264206, 35265645, 35265653
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1492-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 701814RE
- Quantity
- 385 units
- Codes and lots
- UDI/DI 20705032056820; Lot Numbers: 35264219, 35265391, 35265392, 35265399, 35265668, 35265660
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1493-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 601814RM
- Quantity
- 27 units
- Codes and lots
- UDI/DI 20705032056738; Lot Numbers: 35265658, 35264211, 35264225, 35265647
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1494-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 501814RM
- Quantity
- N/A
- Codes and lots
- UDI/DI 20705032054796; Lot Numbers: 35265644
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1495-2023
ANGIOGUARD RX / XP Emboli Capture Guidewire System…
- Product
- ANGIOGUARD RX / XP Emboli Capture Guidewire System, REF 801814RE
- Quantity
- N/A
- Codes and lots
- UDI/DI 20705032056929; Lot Numbers: 35267611
- Recall started
- March 24, 2023
- Device type
- Temporary Carotid Catheter For Embolic Capture
- Root cause (FDA)
- Vendor change control
- What the company told customers
- The firm issued an Urgent MEDICAL DEVICE RECALL notice to it consignees on 03/24/2023 by letter. The notice explained the issue and the hazard, and requested the following actions be taken: - Check inventory and isolate the product to avoid use. - Complete and return the response form - Share this letter with others in your facility who need to be made aware of this recall and with any other facility that may have been sent the affected units of product from your facility. If any units of the affected lots are found to be at the other facility, please arrange the return of the units. - Return any affected product UPDATED 5/15/2023: The firm issued an updated letter dated 4/20/2023 since they extended the scope to include additional item and lot numbers, as well as removing some lot numbers. The remainder of the letter was similar to the initial letter dated 3/24/2023 and also contained a response form for return.
- Recall number
- Z-1496-2023
What should I do? How serious is a recall? Classes and warnings explained