FDA · Medical device recall
Centricity PACS Software Version 7.0 SP0.0.4.7
- Date
- Aug. 3, 2022
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- GE Healthcare, LLC
- Product
- Centricity PACS Software Version 7.0 SP0.0.4.7
- Reason
- The Event Notification Manager (ENM) functionality for certain products does not process notifications for study modifications performed on post verified exams.
- Distribution pattern
- Worldwide distribution - US Nationwide distribution in the states of CO, NV, OH, PA, WA and the countries of Brazil, Canada, China, France, Germany, Israel, Italy, Japan, Oman, Spain, Switzerland, and United Kingdom.
- Quantity
- 47 installations
- Codes and lots
- Model: 5831882-012, Software Version 7.0 SP0.0.4.7 GTIN: 00840682145572
- Recall started
- June 14, 2022
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design
- What the company told customers
- An Urgent Medical Device Correction letter was sent to customers beginning June 14, 2022. The letter informs the customer of the issue and provides safety instructions for continued use of the device. GE Healthcare will correct all affected devices via a software update. In the U.S., this letter will be sent by a traceable means. This letter will be sent to the following titles within the affected accounts: Director/Manager of Radiology, Director/Manager of Cardiology, Risk Manager/Hospital Administrator, Head of Radiology Department, Head of Cardiology Department, PACS Administrator, Director of IT Department, Head of Biomedical Engineering, Head of Imaging Informatics.
- Recall number
- Z-1466-2022
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-1466-2022. The agency record governs; check it and the company before acting.