FDA · Medical device recall

VITROS Chemistry GENT Reagent REF 680 1711…

Sept. 1, 2005

Date posted by the FDA
Sept. 1, 2005
Status
Terminated
Filed by
Ortho-Clinical Diagnostics
Product
VITROS Chemistry GENT Reagent REF 680 1711, *** Responsible firm on label: Ortho-Clinical Diagnostics Inc., Rochester, NY 14626. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests.
Reason
Negatively biased results can be observed when the reagent pack has less than 12 tests remaining and has been on the analyzer greater than 24 hours.
Distribution pattern
The product was shipped nationwide and to Ortho-Clinical/J&J affiliates in Australia, Canada, Singapore, England, France, Germany, Italy and Spain.
Quantity
443 boxes
Codes and lots
01-5846, Exp. 31 May 2006; 01-6036 Exp 31 October 2006; 01-6239 Exp 31 January 2007
Recall started
July 19, 2005
Device type
Enzyme Immunoassay, Gentamicin
Root cause (FDA)
Other
What the company told customers
Letters dated 7/19/2005 with instructions to process QC fluid according to the IFU when processing patient samples. When replacement product arrives, instructions are to destroy the remaining inventory of the recalled lots.
Recall number
Z-1460-05

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-1460-05. The agency record governs; check it and the company before acting.