VITROS Chemistry GENT Reagent REF 680 1711…
- Date posted by the FDA
- Sept. 1, 2005
- Status
- Terminated
- Filed by
- Ortho-Clinical Diagnostics
- Product
- VITROS Chemistry GENT Reagent REF 680 1711, *** Responsible firm on label: Ortho-Clinical Diagnostics Inc., Rochester, NY 14626. Each box contains 6 packs. Each pack contains liquid reagent for 50 tests.
- Reason
- Negatively biased results can be observed when the reagent pack has less than 12 tests remaining and has been on the analyzer greater than 24 hours.
- Distribution pattern
- The product was shipped nationwide and to Ortho-Clinical/J&J affiliates in Australia, Canada, Singapore, England, France, Germany, Italy and Spain.
- Quantity
- 443 boxes
- Codes and lots
- 01-5846, Exp. 31 May 2006; 01-6036 Exp 31 October 2006; 01-6239 Exp 31 January 2007
- Recall started
- July 19, 2005
- Device type
- Enzyme Immunoassay, Gentamicin
- Root cause (FDA)
- Other
- What the company told customers
- Letters dated 7/19/2005 with instructions to process QC fluid according to the IFU when processing patient samples. When replacement product arrives, instructions are to destroy the remaining inventory of the recalled lots.
- Recall number
- Z-1460-05