FDA · Medical device recall
Zeiss brand OPMI Pentero C…
- Date
- June 12, 2013
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Carl Zeiss Meditec, Inc.
- Product
- Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers: 302583-9001-000 (PENTERO C), and 305961-0000-00 (Ceiling Mount for PENTERO C), Product Usage: A surgical microscope and accessories is an AC-powered device intended for use during surgery to provide a magnified view of the surgical field.
- Reason
- Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to have 3 (three) broken screws out of 6 (six) in the lower section of the ceiling mount.
- Distribution pattern
- US Nationwide Distribution.
- Quantity
- 6 units
- Codes and lots
- Serial Numbers: 6725000708, 6725000709, 6725000710, 6725000713, 6725100725, 6725120813
- Recall started
- April 18, 2013
- Device type
- Microscope, Surgical
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Carl Zeiss Meditec, Inc sent a Customer Notification letter to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter stated that firm's sales representative will be scheduling an on-site visit, informing them of the affected product and providing instructions on the recall. For question call 925-580-5371.
- Recall number
- Z-1440-2013
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-1440-2013. The agency record governs; check it and the company before acting.