GE Healthcare Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx and 2 more products

May 6, 2015 Class II

Date
May 6, 2015
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare
Reason
An issue was identified with the mAs accuracy check performed on the system that could result in inaccuracies in the 10% to 20% range for exposure times longer than 15 msec.
Distribution pattern
US Distribution to the states of: AZ, FL, GA, MO and WI.

3 products in this recall

GE Healthcare Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx

Product
GE Healthcare Automatic Mobile X-Ray (AMX) Series: Brivo XR285amx
Quantity
6 units US
Codes and lots
BrivoXR285amx - PN 5555000, 5555000-2
Recall started
Dec. 19, 2014
Device type
System, X-Ray, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare planned action: 1. The customer notification letter will be distributed to the facilities that have installed the following AMX series X-ray systems: Models Brivo XR285 AMX, Optima XR200 AMX, and Optima XR220 AMX. 2. As stated in the customer notification letter, the correction will be provided free of charge and subsequent communication between the firm and the facility will result in a site visit by a GE Healthcare field engineer to bring the X-ray system into compliance. 3. The GEHC field engineer will follow Field Modification Instruction 10899 which provides instructions on how to perform a mAs accuracy check and how to determine if the test results pass predetermined acceptance criteria. If you have any questions or concerns regarding this notification, please call the Information following phone number: 1-800-437-1171.
Recall number
Z-1405-2015

GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR200 amx

Product
GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR200 amx
Quantity
6 units in US
Codes and lots
OptimaXR200amx PN 5555000-3, PN5555000-4
Recall started
Dec. 19, 2014
Device type
System, X-Ray, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare planned action: 1. The customer notification letter will be distributed to the facilities that have installed the following AMX series X-ray systems: Models Brivo XR285 AMX, Optima XR200 AMX, and Optima XR220 AMX. 2. As stated in the customer notification letter, the correction will be provided free of charge and subsequent communication between the firm and the facility will result in a site visit by a GE Healthcare field engineer to bring the X-ray system into compliance. 3. The GEHC field engineer will follow Field Modification Instruction 10899 which provides instructions on how to perform a mAs accuracy check and how to determine if the test results pass predetermined acceptance criteria. If you have any questions or concerns regarding this notification, please call the Information following phone number: 1-800-437-1171.
Recall number
Z-1406-2015

GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR220 amx

Product
GE Healthcare Automatic Mobile X-Ray (AMX) Series: Optima XR220 amx
Quantity
6 units in US
Codes and lots
OptimaXR220amx PN 5555000-5, PN5555000-6
Recall started
Dec. 19, 2014
Device type
System, X-Ray, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare planned action: 1. The customer notification letter will be distributed to the facilities that have installed the following AMX series X-ray systems: Models Brivo XR285 AMX, Optima XR200 AMX, and Optima XR220 AMX. 2. As stated in the customer notification letter, the correction will be provided free of charge and subsequent communication between the firm and the facility will result in a site visit by a GE Healthcare field engineer to bring the X-ray system into compliance. 3. The GEHC field engineer will follow Field Modification Instruction 10899 which provides instructions on how to perform a mAs accuracy check and how to determine if the test results pass predetermined acceptance criteria. If you have any questions or concerns regarding this notification, please call the Information following phone number: 1-800-437-1171.
Recall number
Z-1407-2015

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-1405-2015, Z-1406-2015, Z-1407-2015. The agency record governs; check it and the company before acting.