FDA · Medical device recall
Renessa RF System…
- Date posted by the FDA
- April 16, 2010
- Status
- Terminated
- Filed by
- Novasys Medical Inc
- Product
- Renessa RF System, model PR0918, manufactured by Novasys Medical, Newark, CA
- Reason
- Instructions for Use (IFU) revision-- Because of complaints received by the firm of worsened stress incontinence symptoms 3-4 weeks after treatment, the instructions for use were revised to emphasize potential side effects. Although Novasys has not, to date, identified a device malfunction associated with the symptoms discussed above, the investigation over time shows that there is a correlation
- Distribution pattern
- US nationwide only. (TX, FL, CA, LA, KY, MO, GA,OR WA, HI, AZ, OK AL, TN) .
- Quantity
- As of 6/12/09 there were 436 Probes at customer accounts.
- Codes and lots
- Issue affects all lots in distribution.
- Recall started
- July 27, 2009
- Device type
- Applicator, Transurethral, Radio Frequency, For Stress Urinary Incontinence In Women
- Root cause (FDA)
- Labeling design
- What the company told customers
- A Field Action notification letter was sent to all current customers (physicians) along with the revised instructions for use, and a small supply of the revised patient information brochures. The company also updated physician information brochures and patient/physician DVDs to include the information. This was shipped to physicians in August/September of 2009. Additional copies of the Patient Brochures can be ordered by contacting Novasys Customer Service at 866-784-4777. Questions should be addressed to the firm at the same telephone number.
- Recall number
- Z-1404-2010