FDA · Medical device recall
Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS¿ ECi/ECiQ/3600 Immunodiagnostic…
- Date
- July 20, 2022
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- Ortho-Clinical Diagnostics, Inc.
- Product
- Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS¿ ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS¿ 5600/XT 7600 Integrated Systems. Product Code: 6199960
- Reason
- On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).
- Distribution pattern
- Worldwide distribution - US Nationwide and the countries of Colombia, India, Singapore.
- Quantity
- 20858
- Codes and lots
- UDI-DI :10758750034413. All expired, in-date and future released lots
- Recall started
- May 26, 2022
- Device type
- Reagent, Coronavirus Serological
- Root cause (FDA)
- Incorrect or no expiration date
- What the company told customers
- Ortho Clinical Diagnostics issued Important Correction Notification letter (CL2022-135) on May 26, 2022. Letter states reason for recall, health risk and action to take: When using VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack, do not store on-analyzer for longer than 6 weeks. " Until a lot update occurs, OAS for VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack will temporarily require manual monitoring and removal of reagent packs from analyzer no longer than 6 weeks after packs have been opened and loaded onto the analyzer. " Complete the enclosed Confirmation of Receipt form no later than June 9, 2022. " Post this notification by each VITROS ECi/ECiQ/3600/5600/XT 7600 System in your laboratory until further notice. " Please forward this notification if VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack was distributed outside of your facility. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311
- Recall number
- Z-1367-2022
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-1367-2022. The agency record governs; check it and the company before acting.