FDA · Medical device recall

LotFancy Urinary Tract Infection Urine (UTI) Test Strips… and 10 more products

Feb. 11, 2026 Class II

Date
Feb. 11, 2026
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Changchun Wancheng Bio-Electron Co., Ltd.
Reason
Test strips and other tests including for near-patient testing, that require a 510(k) were distributed before they could be legally marketed to U.S. customers; devices may cause false or inaccurate diagnostic result by lay person and lead to inappropriate medical intervention.
Distribution pattern
US Nationwide distribution in the states of TX, GA, CA.

11 products in this recall

LotFancy Urinary Tract Infection Urine (UTI) Test Strips…

Product
LotFancy Urinary Tract Infection Urine (UTI) Test Strips; PALINOIA UTI-10 PARAMETER REAGENT STRIPS; Vaunn Medical 8-in-1 Urinalysis Reagent Strip; JNW Direct 4-in-1 UTl Urinalysis Test Strips; Urinify STRENGTH; Urinify ESSENTIAL, Urinify Urinary Tract Infection (UTI)
Quantity
91,366
Codes and lots
Name/Lot(Expiration): LotFancy Urinary Tract Infection Urine (UTI) Test Strips/2025062001(06/19/2027); LotFancy URS-3N Test Strip/2025033101(03-30-2027), 2025080801(08-07-2027), 2025090844(09-07-2027); PALINOIA UTI-10 PARAMETER REAGENT STRIPS/2025042110(04-20-2027), 2025021001(02-20-2027); Vaunn Medical 8-in-1 Urinalysis Reagent Strip/2024032508(03/24/2026), 2024071602(07/15/2026), 2024102108(10/20/2026); JNW Direct 4-in-1 UTl Urinalysis Test Strips/30D25WC(04/29/2027), 02D25WC(04/01/2027), 30E25WC(05/29/2027), 10H24WC(07/10/2026); Urinify STRENGTH/202403103-6(01/03/2026), 202504103-6(01/01/2027), 202505103-6(06/01/2027), 202504103-3(01/01/2027), 202505103-3(06/01/2027); Urinify ESSENTIAL/202403104-6(01/03/2026), 202504104-6(01/01/2027), 202505104-6(06/01/2027), 202504104-3(01/01/2027), 202505104-3(06/01/2027); Urinify Urinary Tract Infection (UTI)/202501105-5(01/02/2027)
Recall started
Nov. 22, 2025
Device type
Test, Urine Leukocyte
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1257-2026

The Girlfriend Doctor Dr…

Product
The Girlfriend Doctor Dr. Anna Cabeca Keto-pH-Uric Acid Test Strips; URIMED Urinalysis Reagent Strips URS-5, Rapid UTI Test Includes Ketones: Leukocytes, Nitrite, pH, Ketone, Protein; Vivoo Ketone Test
Quantity
32,500
Codes and lots
Name/Lot(Expiration): The Girlfriend Doctor Dr. Anna Cabeca Keto-pH-Uric Acid Test Strips/2024032508(03/24/2026), 2024071602(07/15/2026), 2024102108(10/20/2026); URIMED Urinalysis Reagent Strips URS-5, Rapid UTI Test Includes Ketones: Leukocytes, Nitrite, pH, Ketone, Protein/2025043007(04/29/2027); Vivoo Ketone Test/2024013122(01/30/2026)
Recall started
Nov. 22, 2025
Device type
Nitroprusside, Ketones (Urinary, Non-Quant.)
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1258-2026

LH One Step Ovulation Test Device (Urine)

Product
LH One Step Ovulation Test Device (Urine)
Quantity
1,000
Codes and lots
Name/Lot(Expiration): LH One Step Ovulation Test Device (Urine)/GT-LH 20251601(05-2027)
Recall started
Nov. 22, 2025
Device type
Test, Luteinizing Hormone (Lh), Over The Counter
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1259-2026

Menopause Test Cassette (Urine)

Product
Menopause Test Cassette (Urine)
Quantity
1,000
Codes and lots
Name/Lot(Expiration): Menopause Test Cassette (Urine)/GT-FSH 20251601(05-2027)
Recall started
Nov. 22, 2025
Device type
Test, Follicle Stimulating Hormone (Fsh), Over The Counter
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1260-2026

S. Typhi/Para Typhi A Antigen

Product
S. Typhi/Para Typhi A Antigen
Quantity
1,000
Codes and lots
Name/Lot(Expiration): S. Typhi/Para Typhi A Antigen/GT-STA 20251601(05-2027)
Recall started
Nov. 22, 2025
Device type
Antiserum, Typhus Fever
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1261-2026

Male Fertility Sperm Test for Home Use (Cassette)

Product
Male Fertility Sperm Test for Home Use (Cassette)
Quantity
1,000
Codes and lots
Name/Lot(Expiration): Male Fertility Sperm Test for Home Use (Cassette)/GT-SPA 20251601(05-2027)
Recall started
Nov. 22, 2025
Device type
Material, Quality Control, Semen Analysis
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1262-2026

Exploro Highly Sensitive Male Fertility / Sperm Concentration Test

Product
Exploro Highly Sensitive Male Fertility / Sperm Concentration Test
Quantity
4,800
Codes and lots
Name/Lot(Expiration): Exploro Highly Sensitive Male Fertility / Sperm Concentration Test/2025033101(03-30-2027), 2025080801(08-07-2027), 2025090844(09-07-2027)
Recall started
Nov. 22, 2025
Device type
Sperm, Antigen, Antiserum, Control
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1263-2026

Vivoo pH Test

Product
Vivoo pH Test
Quantity
500
Codes and lots
Name/Lot(Expiration): Vivoo pH Test/2024013121(01/30/2026)
Recall started
Nov. 22, 2025
Device type
Dye-Indicator, Ph (Urinary, Non-Quantitative)
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1264-2026

Vivoo Protein Test

Product
Vivoo Protein Test
Quantity
500
Codes and lots
Name/Lot(Expiration): Vivoo Protein Test/2024013122(01/30/2026)
Recall started
Nov. 22, 2025
Device type
Indicator Method, Protein Or Albumin (Urinary, Non-Quant.)
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1265-2026

Vivoo Vaginal pH Test

Product
Vivoo Vaginal pH Test
Quantity
3,000
Codes and lots
Name/Lot(Expiration): Vivoo Vaginal pH Test/2024032219(03/21/2026)
Recall started
Nov. 22, 2025
Device type
Paper, Obstetric Ph
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1266-2026

Vivoo Sodium Test…

Product
Vivoo Sodium Test; Vivoo Vitamin C Test; Vivoo Hydration Test
Quantity
3,300
Codes and lots
Name/Lot(Expiration): Vivoo Sodium Test/2024013123(01/30/2026); Vivoo Vitamin C Test/2024013124(01/30/2026); Vivoo Hydration Test/2024013119(01/30/2026), 2025051519(05/14/2027)
Recall started
Nov. 22, 2025
Device type
Urinalysis Controls (Assayed And Unassayed)
Root cause (FDA)
No Marketing Application
What the company told customers
Starting on 11/22/2025, recall notices were mailed or emailed to customers/distributors who were asked to do the following: 1) Destroy unused devices. 2) Inform customers that test results may be inaccurate, and it is recommended to re-test. 3) If affected devices were further distributed, please forward this notice and response form to these additional customers or email us a customer distribution list with names, emails, addresses, phone numbers so we can notify your customers of the recall. 4) Complete and return the Acknowledgement and Receipt Form via email to: sales04@ccwancheng.com Questions can be emailed to the firm at: sales01@ccwancheng.com / sales04@ccwancheng.com
Recall number
Z-1267-2026

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-1257-2026, Z-1258-2026, Z-1259-2026, Z-1260-2026, Z-1261-2026, Z-1262-2026, Z-1263-2026, Z-1264-2026 and others. The agency record governs; check it and the company before acting.