FDA · Medical device recall

MSK Extreme MR Scanner… and 1 more product

May 8, 2013 Class II

Date
May 8, 2013
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Reason
Coolant supply lines may leak liquid onto the floor resulting in a slippery surface and fall hazard.
Distribution pattern
Worldwide Distribution - USA including AL, CA, CO, FL, ID, IL, MA, NH, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI; Worldwide: Argentina, Australia, Austria, Brazil, Canada, Chile, Denmark, Egypt, Finland, Germany, India, Italy, Japan, Korea, Malaysia, Netherlands, Norway, Peru, Poland, Russia, Singapore, South Africa, Spain, Sweeden, Switzerland, UK, Ukraine, and UAE.

2 products in this recall

MSK Extreme MR Scanner…

Product
MSK EXTREME MR SCANNER, MODEL AA5000 The intended for use as a diagnostic imaging device to produce axial, sagittal, coronal and oblique images of the internal structures of the leg (excluding the thigh), knee, ankle, foot, forearm, elbow, wrist and hand.
Quantity
70
Codes and lots
01072011-001; 01082010-001; 01182010-001; 01182011-001; 01272010-001; 02032011-001; 02112010-001; 02182011-001; 02242010-001; 2262011-001; 03012010-001; 03072010-001; 03132009-001; 03242009-001; 03292011-001; 03302011-001; 04012010-001; 04082010-001; 04192011-001; 04202011-001; 04302010-001; 05072010-001; 05132011-001; 05152009-001; 05242010-001; 05262010-001; 06012008-001; 06042009-001; 06062009-001; 06082010-001; 06182010-001; 06192009-001; 07082010-001; 07092010-001; 07142009-001; 07152011-001; 07272010-001; 08082009-001; 08122009-001; 08122010-001; 08132009-001; 08152010-001; 09072009-001; 09132010-001; 09142010-001; 09172009-001; 09222009-001; 09222010-001; 10012009-001; 10082009-001; 10082010-001; 10142009-001; 10182010-001; 10222010-001; 10252009-001; 10312010-001; 11162010-001; 11242009-001; 11292010-001; 12052010-001; 12072009-001; 12142009-001; 12152009-001; 12162008-001; 12182008-001; 12222010-001
Recall started
March 28, 2013
Device type
System, Nuclear Magnetic Resonance Imaging
Root cause (FDA)
Process control
What the company told customers
GE Healthcare sent an Urgent Medical Device Correction letter dated April 23, 2013, to all affected customers. The letter informed customers of a potential safety issue due to leakage from gradient coolant lines. Consignees were instructed to monitor the floor for leakage and continue to use if leakage is not observed. If leakage is observed consignees are instructed to stop using the system and notify their local service representative. GE Healthcare will correct all affected scanners by providing an update at no cost. Customers with questions were instructed to contact their local service representataive. For questions regarding this recall call 262-513-4122.
Recall number
Z-1211-2013

Optima MR430s MRI Scanner…

Product
Optima MR430s MRI Scanner; 5000-0002 is indicated to use as a magnetic resonance imaging device of the leg (excluding the thigh), knee, ankle, foot, elbow, forearm, wrist, and hand.
Quantity
81
Codes and lots
08022011-001; 09092011-001;12102011-001;05152012-004;SM11062012-001;02022012-001;08142012-001;04212012-001;05252012-002;12132011-001;SM08292012-007;11052011-002;01192012-001;01132012-001;06282012-010;07252012-001;01022012-001;11202011-001;08142011-001;06302011-001;09172011-001;10082011-001;10172011-002;03142012-007;06142012-001;06212012-001;07172012-001;07232012-001;SM09132012-003;02042012-001;01132012-001;07172011-001;09062011-001;09242011-001;SM10122012-003;M12102012-003;05102012-004;08022012-003;07302012-008;SM01072013-005;05032012-004;08032012-001;10172011-001;SM08152012-001;12262011-001;10092011-001;04172012-001;11052011-001;07112011-001;10242011-007;07222011-001;08102011-001;03092012-001;05172011-001;06222011-001;03202012-004;05182012-001;07112012-002;06092011-001;01232012-001;02162012-001;11262012-007;SM12172012-001;01212012-001;SM10232012-001;11212011-001;12222011-001;SM11132012-002;06012011-001;07132011-001;12042011-001;03062012-007;03282012-001;10272011-009;12142011-001;10302012-006;02112012-001;03062012-008;01272012-001;05162012-003;03262012-002;11292011-001;06112012-004;10172012-001;12102012-001
Recall started
March 28, 2013
Device type
System, Nuclear Magnetic Resonance Imaging
Root cause (FDA)
Process control
What the company told customers
GE Healthcare sent an Urgent Medical Device Correction letter dated April 23, 2013, to all affected customers. The letter informed customers of a potential safety issue due to leakage from gradient coolant lines. Consignees were instructed to monitor the floor for leakage and continue to use if leakage is not observed. If leakage is observed consignees are instructed to stop using the system and notify their local service representative. GE Healthcare will correct all affected scanners by providing an update at no cost. Customers with questions were instructed to contact their local service representataive. For questions regarding this recall call 262-513-4122.
Recall number
Z-1212-2013

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-1211-2013, Z-1212-2013. The agency record governs; check it and the company before acting.