FDA · Medical device recall
Innova IGS 630… and 2 more products
- Date
- April 24, 2019
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- GE Healthcare, LLC
- Reason
- There is a potential for loss of the x-ray imaging function when the user changes field of view (FOV) from 30cm to 20cm or from 20cm to 30cm while releasing the fluoroscopy footswitch pedal simultaneously.
- Distribution pattern
- Nationwide distribution to AR, CA, CA, FL, MI, NM, NV, SC, TN, TX, VA, WA, WI. International distribution to Algeria, Belgium, China, Colombia, France, Hungary, Italy, Japan, Korea (Republic Of), Kuwait, Libya, Pakistan, Poland, Russia, Saudi Arabia, South Africa, Sudan, Turkey, Vietnam.
3 products in this recall
Innova IGS 630…
- Product
- Innova IGS 630, Angiographic X-Ray, s5i system option used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
- Quantity
- 1 device
- Codes and lots
- Model Number 5503558; System ID Number 0850060757; Serial Number B3-16-015
- Recall started
- March 15, 2019
- Device type
- System, X-Ray, Angiographic
- Root cause (FDA)
- Device Design
- What the company told customers
- GE Healthcare notified customers on about 03/15/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customer were informed that they can continue to use their Innova IGS 6 or Innova IGS 630 Systems but must follow the following safety instructions: 1. Do not change the size of the field of view from 30cm to 20cm or 20cm to 30cm while releasing the fluoroscopy footswitch pedal. 2. Only perform these 2 actions sequentially. 3. If the system becomes locked up, perform a reset to recover functionalities. Customers were also informed that GE Healthcare will correct all affected products at no cost and a GE Healthcare representative will contact each customer to perform the correction. After the GE representative has updated the system, please destroy previous Innova IGS 6 SW package installation media related to this system (versions prior to IGS6_2.2.1 or IGS6_3.1.1). Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Recall number
- Z-1199-2019
Innova IGS 630…
- Product
- Innova IGS 630, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
- Quantity
- 49 devices
- Codes and lots
- Model Number 5730860; System ID (Serial Number, UDI Number): 352333VASC3 (B3-17-006, 01008406821086381117020021B3-17-006); 727869SPECIALS (B3-18-005, 01008406821086381118020021B3-18-005); 407518BIPLANE (B3-17-005, 01008406821086381117010021B3-17-005); 305227KBP630 (B3-16-017, 01008406821086381116080021B3-16-017); 401BIPLANE1 (B3-18-001, 01008406821086381118010021B3-18-001); 941629IGS630 (B3-18-002, 01008406821086381118010021B3-18-002); 505841CATH3 (B3-17-032, 01008406821086381117120021B3-17-032); 7022553131 (B3-17-011, 01008406821086381117030021B3-17-011); 7028803131 (B3-17-023, 01008406821086381117080021B3-17-023); 615769SP3 (B3-17-013, 01008406821086381117040021B3-17-013); 214618INIGS (B3-16-029, 01008406821086381116120021B3-16-029); 281890BP (B3-17-024, 01008406821086381117090021B3-17-024); 956632IGS630 (B3-17-029, 01008406821086381117120021B3-17-029); 804320IGS630 (B3-16-016, 01008406821086381116070021B3-16-016); DZ1009VA03 (B3-17-015, 01008406821086381117070021B3-17-015); 220826RX10 (B3-17-003, 01008406821086381117010021B3-17-003); 082416040099 (B3-19-001, 01008406821086381119020021B3-19-001); 082416160088 (B3-18-023, 01008406821086381118110021B3-18-023); 082416100170 (B3-19-003, 01008406821086381119020021B3-19-003); XV8624CO (B3-18-012, 01008406821086381118050021B3-18-012); N4192956 (B3-17-009, 01008406821086381117030021B3-17-009); A5820150 (B3-16-023, 01008406821086381116100021B3-16-023); YV1920 (B3-16-020, 01008406821086381116100021B3-16-020); YV1922 (B3-17-002, 01008406821086381117010021B3-17-002); YV1921 (B3-16-027, 01008406821086381116120021B3-16-027); YV4500 (B3-16-032, 01008406821086381117010021B3-16-032); YV4501 (B3-17-027, 01008406821086381117110021B3-17-027); YV2503 (B3-16-019, 01008406821086381116090021B3-16-019); YV4503 (B3-18-003, 01008406821086381118020021B3-18-003); YV4502 (B3-17-030, 01008406821086381117120021B3-17-030); 0850060758 (B3-17-014, 01008406821086381117070021B3-17-014); 0850060755 (B3-17-001, 01008406821086381117010021B3-17-001); 0850060759 (B3-18-013, 01008406821086381118060021B3-18-013); KW1008VA01 (B3-17-019, 01008406821086381117070021B3-17-019); KW1005VA02 (B3-16-025, 01008406821086381116110021B3-16-025); KW1005VA01 (B3-16-026, 01008406821086381116110021B3-16-026); LY1067VA01 (B3-17-010, 01008406821086381117030021B3-17-010); PK1003VA01 (B3-17-018, 01008406821086381117060021B3-17-018); RU2011VA02 (B3-17-004, 01008406821086381117010021B3-17-004); RU9774VA01 (B3-16-021, 01008406821086381116100021B3-16-021); SA2513VA01 (B3-16-024, 01008406821086381116100021B3-16-024); ZA2940VA01 (B3-17-025, 01008406821086381117100021B3-17-025); SD1084VA01 (B3-18-010, 01008406821086381118050021B3-18-010); 06189VAS06 (B3-17-028, 01008406821086381117110021B3-17-028); 55001VAS02 (B3-16-031, 01008406821086381116120021B3-16-031); 5973XR0085 (B3-17-007, 01008406821086381117020021B3-17-007); 5973XR0075 (B3-16-030, 01008406821086381116120021B3-16-030); 5973XR0076 (B3-17-008, 01008406821086381117030021B3-17-008); 5973XR0094 (B3-17-012, 01008406821086381117040021B3-17-012)
- Recall started
- March 15, 2019
- Device type
- System, X-Ray, Angiographic
- Root cause (FDA)
- Device Design
- What the company told customers
- GE Healthcare notified customers on about 03/15/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customer were informed that they can continue to use their Innova IGS 6 or Innova IGS 630 Systems but must follow the following safety instructions: 1. Do not change the size of the field of view from 30cm to 20cm or 20cm to 30cm while releasing the fluoroscopy footswitch pedal. 2. Only perform these 2 actions sequentially. 3. If the system becomes locked up, perform a reset to recover functionalities. Customers were also informed that GE Healthcare will correct all affected products at no cost and a GE Healthcare representative will contact each customer to perform the correction. After the GE representative has updated the system, please destroy previous Innova IGS 6 SW package installation media related to this system (versions prior to IGS6_2.2.1 or IGS6_3.1.1). Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Recall number
- Z-1200-2019
Innova IGS 6…
- Product
- Innova IGS 6, Interventional Fluoroscopic X-Ray System used in generating fluoroscopic and rotational images of human anatomy for cardiovascular, vascular and non-vascular, diagnostic and interventional procedures.
- Quantity
- 22 devices
- Codes and lots
- Model Number 5771387; System ID (Serial Number, UDI Number): 501257IGS630 (B3-18-014, 01008406821246141118070021B3-18-014); 951486RU630 (B3-18-009, 01008406821246141118040021B3-18-009); 714993IGS630 (B3-18-004, 01008406821246141118020021B3-18-004); 906225EP1 (B3-18-015, 01008406821246141118070021B3-18-015); 906225CATH4 (B3-18-016, 01008406821246141118080021B3-18-016); 843797IGS30 (B3-18-030, 01008406821246141119010021B3-18-030); 4603984XV41 (B3-18-019, 01008406821246141118090021B3-18-019); MCARIGS630 (B3-18-008, 01008406821246141118040021B3-18-008); 509473RM1B (B3-18-028, 01008406821246141119010021B3-18-028); 414649IGS8 (B3-18-020, 01008406821246141118090021B3-18-020); HU1099VA07 (B3-18-024, 01008406821246141118110021B3-18-024); A670060221 (B3-19-002, 01008406821246141119020021B3-19-002); A5163326 (B3-18-007, 01008406821246141118030021B3-18-007); YV4504 (B3-18-018, 01008406821246141118090021B3-18-018); YV4505 (B3-18-027, 01008406821246141118120021B3-18-027); YV4507 (B3-18-022, 01008406821246141118110021B3-18-022); 0850060762 (B3-18-025, 01008406821246141118110021B3-18-025); 0850060763 (B3-18-026, 01008406821246141118120021B3-18-026); KW1155VA02 (B3-18-017, 01008406821246141118080021B3-18-017); PL1689VA01 (B3-18-021, 01008406821246141118100021B3-18-021); PL0302VA02 (B3-18-011, 01008406821246141118060021B3-18-011); 06189VAS11 (B3-18-029, 01008406821246141119010021B3-18-029)
- Recall started
- March 15, 2019
- Device type
- System, X-Ray, Angiographic
- Root cause (FDA)
- Device Design
- What the company told customers
- GE Healthcare notified customers on about 03/15/2019 via "URGENT MEDICAL DEVICE CORRECTION" letter. Customer were informed that they can continue to use their Innova IGS 6 or Innova IGS 630 Systems but must follow the following safety instructions: 1. Do not change the size of the field of view from 30cm to 20cm or 20cm to 30cm while releasing the fluoroscopy footswitch pedal. 2. Only perform these 2 actions sequentially. 3. If the system becomes locked up, perform a reset to recover functionalities. Customers were also informed that GE Healthcare will correct all affected products at no cost and a GE Healthcare representative will contact each customer to perform the correction. After the GE representative has updated the system, please destroy previous Innova IGS 6 SW package installation media related to this system (versions prior to IGS6_2.2.1 or IGS6_3.1.1). Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Recall number
- Z-1201-2019