FDA · Medical device recall
Newport HT50 Ventilator (all Models)
- Date posted by the FDA
- Sept. 19, 2007
- Status
- Terminated
- Filed by
- Newport Medical Instruments Inc
- Product
- Newport HT50 Ventilator (all Models)
- Reason
- Battery Life: With use over time, the internal battery begins losing its ability to hold a charge resulting in shortened ventilator operation time. Specifically, Newport has received a number of reports concerning the internal battery not lasting as long as expected followed by ventilator shut down with less advance warning than when the battery was new.
- Distribution pattern
- Worldwide including USA, Algeria, Argentina, Armenia, Australia, Belgium, Bolivia, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cote D' Ivoire, Croatia, Cyprus, Dominican Republic, Ecuador, Egypt, El Salvador, England, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kenya, Korea, Kuwait, Liechtenstein, Lithuania, Malaysia, Mexico, Netherlands, New Caldonia, New Zealand, Oman, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Switzerland, Syria, Taiwan, Thailand, Turkey, UAE, Ukraine, United Kingdom, Venuzuela, Yemen, and Zambia
- Quantity
- 8,160
- Codes and lots
- All models, all serial numbers
- Recall started
- July 31, 2007
- Device type
- Ventilator, Continuous, Facility Use
- Root cause (FDA)
- Other
- What the company told customers
- An Important Medical Device Correction Letter was issued on September 14, 2007 as a short term fix. The firm has submitted a long term correction as a supplement to its 510 K which is pending review by FDA.
- Recall number
- Z-1192-2007
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-1192-2007. The agency record governs; check it and the company before acting.