FDA · Medical device recall
GE Healthcare Fluorostar 7900 Mobile Fluoroscopy System
- Date posted by the FDA
- Sept. 24, 2009
- Status
- Terminated
- Filed by
- Ge Healthcare
- Product
- GE Healthcare Fluorostar 7900 Mobile Fluoroscopy System
- Reason
- The alarm timer is reset after each exposure and therefore never reaches the five minute alarm state unless one exposure exceeds 5 minutes in duration.
- Distribution pattern
- Worldwide Distribution
- Quantity
- 554 units
- Codes and lots
- n/a
- Recall started
- Feb. 19, 2009
- Device type
- System, X-Ray, Fluoroscopic, Image-Intensified
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- GE Healthcare sent a customer notification out to all sites that have potentially affected systems. GEHC Surgery will also install a software correction into the affected systems in order to resolve this issue at no charge to the customer. Until the GEHC OEC Fluorostar 7900 has received the corrective software installed, they are encouraging facilities to exercise caution in monitoring the amount of cumulative fluoroscopy time when the affected system is in use.
- Recall number
- Z-1155-2009