FDA · Medical device recall

GE Healthcare Fluorostar 7900 Mobile Fluoroscopy System

Sept. 24, 2009

Date posted by the FDA
Sept. 24, 2009
Status
Terminated
Filed by
Ge Healthcare
Product
GE Healthcare Fluorostar 7900 Mobile Fluoroscopy System
Reason
The alarm timer is reset after each exposure and therefore never reaches the five minute alarm state unless one exposure exceeds 5 minutes in duration.
Distribution pattern
Worldwide Distribution
Quantity
554 units
Codes and lots
n/a
Recall started
Feb. 19, 2009
Device type
System, X-Ray, Fluoroscopic, Image-Intensified
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare sent a customer notification out to all sites that have potentially affected systems. GEHC Surgery will also install a software correction into the affected systems in order to resolve this issue at no charge to the customer. Until the GEHC OEC Fluorostar 7900 has received the corrective software installed, they are encouraging facilities to exercise caution in monitoring the amount of cumulative fluoroscopy time when the affected system is in use.
Recall number
Z-1155-2009