Menu
FDA · Medical device recall

Treatment packs used on the VisuMax Laser Keratome…

March 19, 2014 Class II

Date
March 19, 2014
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Carl Zeiss Meditec AG
Product
Treatment packs used on the VisuMax Laser Keratome. Size M Manufactured by Carl Zeiss Meditec AG, Jena, Germany. The VisuMax Laser Keratome is indicated for the following: In the creation of a corneal flap in patients undergoing LASIK surgery or other treatment requiring lamellar resection of the cornea. Patients undergoing surgery or other treatment requiring initial lamellar resection of the cornea. In the creation of a lamellar cut/resection of the cornea for lamellar keratoplasty. In the creation of a cut/incision for penetrating keratoplasty and corneal harvesting.
Reason
Carl Zeiss Treatment pack, lot number M 130010, size M, may actually contain size S.
Distribution pattern
US Distribution: NV only.
Quantity
1 package containing 10 individual treatment packs
Codes and lots
P/N 1462-333, Size M, Lot # M130010
Recall started
Feb. 3, 2014
Device type
Laser, Ophthalmic
Root cause (FDA)
Mixed-up of materials/components
What the company told customers
The firm, Zeiss, sent an "Important Field Safety Advisory Note" letter dated February 3, 2014 to its customer. Letter indicates that a phone call preceded the letter on January 23, 2014. The letter also describes the product, problem and actions to be taken. The customers were instructed to quarantine all treatment packs size "M", batch number M 130010; return the material to Carl Zeiss Meditec, Inc., 5160 Hacienda Dr, Dublin, CA 94568 and Zeiss will issue a credit for all material received, upon receipt of the returned materia, and to complete and return the Conformation Form with the envelop provided. If you have any questions regarding this field corrective action, please call 925-557-4832 or email at paul.shahan@zeiss.com.
Recall number
Z-1153-2014

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-1153-2014. The agency record governs; check it and the company before acting.

Report a mistake on this page