FDA · Medical device recall

VITROS Chemistry Products PHBR Slides…

Feb. 28, 2012

Date posted by the FDA
Feb. 28, 2012
Status
Terminated
Filed by
Ortho-Clinical Diagnostics
Product
VITROS Chemistry Products PHBR Slides, REF 822 1384 --- COMMON/USUAL NAME: VITROS PHBR (Phenobarbital) Slides --- Ortho-Clinical Diagnostics, Inc. Rochester, NY --- Ortho-Clinical Diagnostics High Wycombe, UK --- This is an in vitro diagnostic. For in vitro diagnostic use only. VITROS Chemistry Products PHBR Slides quantitatively measure Phenobarbital (PHBR) concentration in serum and plasma using VITROS 250/350/950 and 5,1 FS Chemistry Systems and the VITROS 5600 Integrated System.
Reason
Ortho Clinical Diagnostics (OCD) received customer complaints of intermittent imprecision, outliers, or shifts in control fluid values when using various lots of VITROS Chemistry Products PHBR Slides (Phenobarbital) from Coating 0053.
Distribution pattern
Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Quantity
Domestic - 2345 sale units, Foreign - 632 sales units
Codes and lots
Lot numbers: 2532-0053-4961 (exp. 2/01/12), 2532-0053-4984 (exp. 2/01/12), 2532-0053-4996 (exp. 3/01/12), 2532-0053-4997 (exp. 3/01/12), 2532-0053-0144 (exp. 3/01/12), 2532-0053-0149 (exp. 3/01/12), 2532-0053-0217 (exp. 3/01/12), 2532-0053-0615 (exp. 4/01/12), 2532-0053-0616 (exp. 4/01/12), 2532-0053-1472 (exp. 5/01/12), 2532-0053-1488 (exp. 5/01/12), 2532-0053-1489 (exp. 5/01/12), 2532-0053-1490 (exp. 5/01/12) [Note the four middle numbers (0053) of the above Lot #s constitute the Coating #].
Recall started
June 23, 2011
Device type
Enzyme Immunoassay, Phenobarbital
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Ortho Clinical Diagnostics sent an "URGENT PRODUCT CORRECTION NOTIFICATION" letter dated June 23, 2011 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructs customers to discontinue using and discard all remaining lots of the affected product. A Confirmation of Receipt Form was attached so customers may complete and return to the firm. The firm recommends that customers consult with their Laboratory Medical Director and requesting physician to resolve any concerns they may have regarding previously reported results. Contact Customer Technical Services at 1-800-421-3311 for questions regarding this notification.
Recall number
Z-1117-2012

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-1117-2012. The agency record governs; check it and the company before acting.