FDA · Medical device recall
Trilogy 100…
- Date
- March 16, 2016
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Philips Respironics
- Product
- Trilogy 100, Trilogy 200, Trilogy O2, Trilogy 202, Trilogy EC, Garbin, Garbin Plus Trilogy 100, Trilogy 200, Garbin, Garbin plus provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
- Reason
- Software Issue
- Distribution pattern
- Nationwide Distribution including Puerto Rico and Internationally to the US Virgin Islands.
- Quantity
- 104,508 units
- Codes and lots
- 11 digit S/N begginning with GV or TV; also TGY1D01463
- Recall started
- Jan. 18, 2016
- Device type
- Ventilator, Continuous, Facility Use
- Root cause (FDA)
- Software design
- What the company told customers
- Philips Respironics has determined that affected devices may continue to be used in accordance with the device Instructions for Use and the Field Safety Notification that is being issued to customers. Philips Respironics will be providing a software update for affected devices free of charge. Customers will be notified by mail when the software update is available for download. For further questions please call ( 724 ) 387-7500.
- Recall number
- Z-1065-2016
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-1065-2016. The agency record governs; check it and the company before acting.