FDA · Medical device recall

Trilogy 100…

March 16, 2016 Class II

Date
March 16, 2016
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Philips Respironics
Product
Trilogy 100, Trilogy 200, Trilogy O2, Trilogy 202, Trilogy EC, Garbin, Garbin Plus Trilogy 100, Trilogy 200, Garbin, Garbin plus provides continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
Reason
Software Issue
Distribution pattern
Nationwide Distribution including Puerto Rico and Internationally to the US Virgin Islands.
Quantity
104,508 units
Codes and lots
11 digit S/N begginning with GV or TV; also TGY1D01463
Recall started
Jan. 18, 2016
Device type
Ventilator, Continuous, Facility Use
Root cause (FDA)
Software design
What the company told customers
Philips Respironics has determined that affected devices may continue to be used in accordance with the device Instructions for Use and the Field Safety Notification that is being issued to customers. Philips Respironics will be providing a software update for affected devices free of charge. Customers will be notified by mail when the software update is available for download. For further questions please call ( 724 ) 387-7500.
Recall number
Z-1065-2016

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-1065-2016. The agency record governs; check it and the company before acting.