FDA · Medical device recall
ZEISS miLOOP Lens Fragmentation Device REF 303071-9090-000…
- Date
- May 4, 2022
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- Carl Zeiss Meditec AG
- Product
- ZEISS miLOOP Lens Fragmentation Device REF 303071-9090-000. Used during cataract surgery.
- Reason
- High friction of the slider can cause the device to stick, or not move as intended.
- Distribution pattern
- U.S.: AL, AR, AZ, CA, CT, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OR, PA, Puerto Rico, SC, TN, TX. UT, VA, WA, and WV O.U.S.: Not provided
- Quantity
- 1225 devices
- Codes and lots
- Model: 303071-9090-000 Catalog Number: FG-50608 UDI: (01)04049539104496(11)201102(17)231031(10)FG21082410 Lot Number: FG21082410
- Recall started
- March 16, 2022
- Device type
- Hook, Ophthalmic
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On 03/16/2022, the Recalling firm called U.S. Customers to inform them of that is has been determined that single lot (Lot FG21082410) of the ZEISS miLoop may have been provided with potentially too high friction of the slider. A follow-up email was sent on 03/22/2022 that contained an "Urgent Medical Device Recall" letter. Customer are asked to discontinue the use and not to further distribute the specific product lot, and to fill out the attached acknowledgement form to dl.med-qualitycti.us@zeiss.com. Customer are asked to return all devices from lot FG21082410 to Carl Zeiss Meditec AG. For question or assistance, contact: Scott Smith, Director of Sales at (704) 906-3287 or email scott.smith@zeiss.com
- Recall number
- Z-1007-2022