Menu
FDA · Medical device recall

Brand Name: VITROS Product Name: VITROS 5600 Integrated System… and 2 more products

Jan. 29, 2025 Class II

Date
Jan. 29, 2025
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Ortho-Clinical Diagnostics, INc.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.

3 products in this recall

Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New)…

Product
Brand Name: VITROS Product Name: VITROS 5600 Integrated System Model/Catalog Number: 6802413 (New), 6802415 (Certified/Refurbished), 6904001 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: No, not a component
Reason
The VITROS 5600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.
Quantity
3,249 units (905 US, 2,344 OUS)
Codes and lots
Lot Code: VITROS 5600 Integrated System (New) Product Code: 6802413 UDI: 10758750002740 VITROS 5600 Integrated System (Certified/Refurbished) Product Code: 6802915 UDI: 10758750002740 Lot Code: VITROS 5600 Integrated System (China) Product Code: 6904001 UDI: 10758750035786 Range of installed serial numbers: 56000015-56004962 Manufacture date range: August 2008 to present Distribution date range; August 2008 to present Total systems currently installed in the US: 905 Total systems currently installed outside the US:2344 Note1: The catalogue number 6802413 is used to denote the VITROS 5600 Integrated System. Catalogue number 6802415 is used to denote those VITROS 5600 Integrated Systems that have been refurbished, these systems can also be referred to as certified. A VITROS 5600 Integrated System maintains its original serial number through the refurbishment/certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today. Note3: The value given for "Product Quantity Distributed" is the total number of active installed systems that are in use today. This number is known to fluctuate as individual systems are installed and/or decommissioned.
Recall started
Dec. 19, 2024
Device type
Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Root cause (FDA)
Software Design Change
What the company told customers
On December 19, 2024 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. REQUIRED ACTION " Do not use expired Signal Reagent packs. See the Questions & Answers section for more details. " Complete the enclosed Confirmation of Receipt form no later than December 31, 2024. " Save this notification with your User Documentation or post this notification by each VITROS 3600/5600/XT 7600 System in your laboratory until the issue has been resolved. Contact Information We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact Global Services Organization at 1-800-421-3311. To report adverse events, contact your local Global Services Organization. Adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA's MedWatch Adverse Event Reporting program: www.fda.gov/medwatch/report.htm, 1-800-332-1088.
Recall number
Z-1005-2025

Brand Name: VITROS Product Name: VITROS XT 7600 Integrated System Model/Catalog Number: 6844461 (New)…

Product
Brand Name: VITROS Product Name: VITROS XT 7600 Integrated System Model/Catalog Number: 6844461 (New), 6272222 (Certified), 6904003 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: No, not a component
Reason
The VITROS XT 7600 Integrated System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.
Quantity
1,893 units (994 US, 899 OUS)
Codes and lots
Lot Code: VITROS XT 7600 Integrated System Product Code: 6844461 (New) UDI: 10758750012343 VITROS XT 7600 System (Certified) Product Code: 6272222 UDI: 10758750012343 Lot Code: VITROS XT 7600 Integrated System Product Code: 6904003 (China) UDI: 10758750035809 Range of installed serial numbers:76000001-76002199 Manufacture date range: July 2018 to present Distribution date range; July 2018 to present Total systems currently installed in the US: 994 Total systems currently installed outside the US: 899 Note1: The catalogue number 6844461 is used to denote the VITROS XT 7600 Integrated System. Catalogue number 6272222 is used to denote VITROS XT 7600 System Certified. A VITROS XT 7600 Integrated System maintains its original serial number through the certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today. Note3: The value given for "Product Quantity Distributed" is the total number of active installed systems that are in use today. This number is known to fluctuate as individual systems are installed and/or decommissioned.
Recall started
Dec. 19, 2024
Device type
Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Root cause (FDA)
Software Design Change
What the company told customers
On December 19, 2024 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. REQUIRED ACTION " Do not use expired Signal Reagent packs. See the Questions & Answers section for more details. " Complete the enclosed Confirmation of Receipt form no later than December 31, 2024. " Save this notification with your User Documentation or post this notification by each VITROS 3600/5600/XT 7600 System in your laboratory until the issue has been resolved. Contact Information We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact Global Services Organization at 1-800-421-3311. To report adverse events, contact your local Global Services Organization. Adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA's MedWatch Adverse Event Reporting program: www.fda.gov/medwatch/report.htm, 1-800-332-1088.
Recall number
Z-1006-2025

Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Model/Catalog Number: 6802783 (New)…

Product
Brand Name: VITROS Product Name: VITROS 3600 Immunodiagnostic System Model/Catalog Number: 6802783 (New), 6802914 (Certified/Refurbished), 6904000 (China) Software Version: version 3.8.0 or higher Product Description: In vitro diagnostic chemistry analyzer. FDA 3 letter product code: JJE Component: no, not a component
Reason
The VITROS 3600 Immunodiagnostic System software versions 3.8.0 may not correctly notify users if the Signal Reagent Pack exceeds the on-board and/or shelf-life expiration date. This could result in the use of expired products, which would increase the risk of no results or erroneous results from the instrument.
Quantity
981 units (99 US, 882 OUS)
Codes and lots
Lot Code: VITROS 3600 Immunodiagnostic System Product Code: 6802783 (New) UDI: 10758750002979 VITROS 3600 Immunodiagnostic System (Certified/Refurbished) Product Code: 6802914 UDI: 10758750002979 VITROS 3600 Immunodiagnostic System (China) Product Code: 6904000 UDI: 10758750035779 Range of installed serial numbers:36000004-36001776 Manufacture date range: June 2008 to present Distribution date range; June 2008 to present Total systems currently installed in the US: 99 Total systems currently installed outside the US: 882 Note1: The catalogue number 6802783 is used to denote the VITROS 3600 Immunodiagnostic System. Catalogue number 6802914 is used to denote the VITROS 3600 Immunodiagnostic System that have been refurbished, these systems can also be referred to as certified. A VITROS 3600 Immunodiagnostic System Certified/Refurbished maintains its original serial number through the refurbishment/certification process. Therefore, the manufacturing and distribution information for the two catalogue numbers will be combined. Note2: When manufactured the system s serial numbers are sequential, but not all analyzers are released for distribution and therefore not installed. The range of serial numbers provided includes the identified minimum and maximum in serial numbers associated with all currently installed systems. Therefore, the amount of each system manufactured may not equal the amount of each system installed. Additionally, not all serial numbers listed are associated with an active installed analyzer in use today. Note3: The value given for "Product Quantity Distributed" is the total number of active installed systems that are in use today. This number is known to fluctuate as individual systems are installed and/or decommissioned.
Recall started
Dec. 19, 2024
Device type
Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Root cause (FDA)
Software Design Change
What the company told customers
On December 19, 2024 URGENT PRODUCT CORRECTION NOTIFICATION letters were sent to customers. REQUIRED ACTION " Do not use expired Signal Reagent packs. See the Questions & Answers section for more details. " Complete the enclosed Confirmation of Receipt form no later than December 31, 2024. " Save this notification with your User Documentation or post this notification by each VITROS 3600/5600/XT 7600 System in your laboratory until the issue has been resolved. Contact Information We apologize for the inconvenience this will cause your laboratory. If you have further questions, please contact Global Services Organization at 1-800-421-3311. To report adverse events, contact your local Global Services Organization. Adverse reactions or quality problems experienced with the use of this product may also be reported to the FDA's MedWatch Adverse Event Reporting program: www.fda.gov/medwatch/report.htm, 1-800-332-1088.
Recall number
Z-1007-2025

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-1005-2025, Z-1006-2025, Z-1007-2025. The agency record governs; check it and the company before acting.

Report a mistake on this page