FDA · Medical device recall
VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemistry Systems (VITROS 5,1 FS…
- Date
- Jan. 28, 2015
- Risk class
- Class III (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences)
- Status
- Terminated
- Filed by
- Ortho-Clinical Diagnostics
- Product
- VITROS Software Version 3.1 utilized on the following systems: 1) 4600 Chemistry Systems (VITROS 5,1 FS System family member, 2) 5600 Integrated System 1) VITROS 4600 - in vitro quantitative measurement of a variety of analytes, 2) VITROS 5600 - in vitro quantitative, semi-quantitative, and qualitative measurement of analytes
- Reason
- Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2.
- Distribution pattern
- Worldwide Distribution - USA including CO, NC, NJ, NY, PA, MI, WV and Internationally to Canada, Japan and Spain.
- Quantity
- 12 units
- Codes and lots
- System Product Code: 1) 6802445, 2) 6802413
- Recall started
- Sept. 29, 2014
- Device type
- Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Root cause (FDA)
- Software Manufacturing/Software Deployment
- What the company told customers
- Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated September 29, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2. The resolution to this software anomaly will be in a future version of software. Until the new software is installed on their VITROS System, customers were instructed not to use the calibrator barcode labels for VITROS Calibrator Kit 2. Customers were advised that you manually program the calibration. Customers with questions were instructed to contact Customer Technical Services at 1-800-421-3311.
- Recall number
- Z-1004-2015