VITROS Chemistry Products Na+ Slides…
- Date
- March 20, 2019
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Ortho-Clinical Diagnostics
- Product
- VITROS Chemistry Products Na+ Slides, Catalog Number 8379034 For in vitro diagnostic use only. VITROS Na+ Slides quantitatively measure sodium (Na+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
- Reason
- Imprecision in the affected lot may produce negatively or positively biased patient sample and quality control fluid results.
- Distribution pattern
- US Distribution to states of: AL, CA, CT, FL, HI, IA, ID, IL, LA, MI, MO, MS, NC, NM, NY, OK, PA, TX, WV, and WY; and internationally to: Canada.
- Quantity
- 1448
- Codes and lots
- Lot 4221-0998-3223
- Recall started
- Jan. 22, 2019
- Device type
- Electrode, Ion Specific, Sodium
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- The firm, Ortho Clinical Diagnostics, sent an "URGENT PRODUCT CORRECTION NOTIFICATION" letters dated 1/22/19 to is customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: " If you have an alternate lot of VITROS Na+ Slides, discontinue using and discard all remaining inventory of Lot 4221-0998-3223. " If you do NOT have an alternate lot, consult with your medical or laboratory director to determine if it is acceptable to continue using the affected lot until your replacement order arrives. " Until you receive your replacement order: o Performing Osmolality calculations on your VITROS System is not recommended. o If you continue using the affected lot, it is strongly recommended to take additional action to detect imprecision, such as performing duplicate testing for each sample. Repeat testing if there is a discrepancy between the results obtained in duplicate testing (i.e., +/- 1.4 mmol/L). o Discard all remaining inventory when your replacement order arrives. An adequate supply of replacement product is available and Ortho will expedite all replacement orders. " Please complete the enclosed Confirmation of Receipt form by e-Mail to RA-OCDUS-CONFIRMAD@ORTHOCLINICALDIAGNOSTICS.COM or Fax to: 1.888.557.3759 or 1.585.453.4110. Ortho will credit your account or send a replacement order for the discarded slides. " Forward the information in this notification if you have distributed this product outside of your facility. If you have questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
- Recall number
- Z-0992-2019