FDA · Medical device recall
MORPH ACCESSPRO Steerable Introducer…
- Date
- Feb. 26, 2014
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Biocardia, Inc.
- Product
- MORPH ACCESSPRO Steerable Introducer; 6F introducer, 90 cm working length: Sterile in unopened, undamaged pouch. Sterilized by EtO gas, Non-pyrogenic: BioCardia Inc. San Carlos, CA 94070 Intended to provide a pathway through which medical instruments such as balloon dilation catheters, guidewires and other therapeutic devices may be introduced into the peripheral vasculature.
- Reason
- Tyvek packaging may be worn, causing loss of package integrity.
- Distribution pattern
- US Distribution including the states of FL, MA, WY. OH, MI and KS.
- Quantity
- 11 units
- Codes and lots
- Catalog number MAP690; Lot numbers 01162 and 01186.
- Recall started
- Feb. 7, 2014
- Device type
- Introducer, Catheter
- Root cause (FDA)
- Process control
- What the company told customers
- Sales Representative called each affected hospital on February 7, 2014, to notify them of the recall. BioCardia has initiated a recall of Morph AccessPro due to a potential compromise of the sterile barrier. Use of this product may result in patient infection. Sales representatives contacted hospital sites identified in Attachment 3 by telephone. Script provided as follows: "BioCardia has initiated a recall of Morph Access Pro due to a potential compromise of the sterile barrier. Use of this product may result in patient infection. Our records indicate that we sold you. The recall is limited to these units in this lot sold to you. We would like to confirm if you still have the unit(s). If you do, we ask that you quarantine the unit(s) and our customer service department will contact you to coordinate the return of the unit(s)." For further questions please call (650) 226-0133.
- Recall number
- Z-0991-2014
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0991-2014. The agency record governs; check it and the company before acting.