FDA · Medical device recall

Medtronic Sofamor Danek brand METRx System 16 mm Med Procedure Kit… and 2 more products

July 13, 2005

Date posted by the FDA
July 13, 2005
Status
Terminated
Filed by
Warsaw Orthopedic, Inc.
Reason
The package is labeled as sterile, but there may be a hole in the immediate package caused by the carton design.
Distribution pattern
Nationwide, Australia, Brazil, Canada, Chile, China, Dubai, Hong Kong, India, Korea, Mexico, Netherlands, South Africa and Taiwan.

3 products in this recall

Medtronic Sofamor Danek brand METRx System 16 mm Med Procedure Kit…

Product
Medtronic Sofamor Danek brand METRx System 16 mm Med Procedure Kit; Ref 9560016.
Codes and lots
All lots.
Recall started
June 20, 2005
Device type
Arthroscope
Root cause (FDA)
Package design/selection
What the company told customers
Consignees were notified, via letter dated June 20, 2005, of the problem and to return all non-expired product.
Recall number
Z-0991-05

Medtronic Sofamor Danek brand METRx System 18 mm Med Procedure Kit…

Product
Medtronic Sofamor Danek brand METRx System 18 mm Med Procedure Kit; Ref 9560018.
Codes and lots
All lots.
Recall started
June 20, 2005
Device type
Arthroscope
Root cause (FDA)
Package design/selection
What the company told customers
Consignees were notified, via letter dated June 20, 2005, of the problem and to return all non-expired product.
Recall number
Z-0992-05

Medtronic Sofamor Danek brand METRx System MD Procedure Kit…

Product
Medtronic Sofamor Danek brand METRx System MD Procedure Kit; Ref 9560030.
Codes and lots
All lots.
Recall started
June 20, 2005
Device type
Arthroscope
Root cause (FDA)
Package design/selection
What the company told customers
Consignees were notified, via letter dated June 20, 2005, of the problem and to return all non-expired product.
Recall number
Z-0993-05

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0991-05, Z-0992-05, Z-0993-05. The agency record governs; check it and the company before acting.