Menu
FDA · Medical device recall

VITROS Chemistry Products Calibrator Kit 24…

Feb. 7, 2012

Date posted by the FDA
Feb. 7, 2012
Status
Terminated
Filed by
Ortho-Clinical Diagnostics
Product
VITROS Chemistry Products Calibrator Kit 24, Catalog No. 680 1708. Common/Usual Name: VITROS Calibrator Kit 24. For in vitro diagnostics use only. VITROS Chemistry Products Calibrator Kit 24 is used to calibrate the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of albumin in urine. Ortho-Clinical Diagnostics Inc.
Reason
Ortho Clinical Diagnostics received numerous complaints of lower than expected results for control fluids and patient samples upon initial calibration or following a recalibration using VITROS Chemistry Products Calibrator Kit 24 and VITROS Chemistry Products mALB Reagent.
Distribution pattern
Worldwide Distribution - USA, including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain, and Venezuela.
Quantity
4436 units worldwide (domestic - 2201; foreign - 2235)
Codes and lots
Kit Lot 2430 (exp. 11/16/11), Kit Lot 2440 (exp. 01/08/12), Kit Lot 2450 (03/09/12), Kit Lot 2471 (exp. 7/14/12)
Recall started
Oct. 27, 2011
Device type
Calibrator, Secondary
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
All consignees were sent "Important Product Correction Notification" communications (dated 10/27/11) beginning on 10/27/11 informing them of the issue, and instructing them to discontinue use of their remaining inventory of VITROS Chemistry Products Calibrator Kit 24, VITROS Chemistry Products mALB reagent, and VITROS Chemistry Products mALB Performance Verifiers. For questions, please contact OCD's Customer Technical Services at 1-800-421-3311.
Recall number
Z-0971-2012

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0971-2012. The agency record governs; check it and the company before acting.

Report a mistake on this page