FDA · Medical device recall
Supergrade Earloop Masks…
- Date posted by the FDA
- June 21, 2003
- Status
- Terminated
- Filed by
- Safco Dental Supply Co. Inc.
- Product
- Supergrade Earloop Masks, surgical mask; 50 masks per box; Distributed by Safco Dental Supply Co., Northbrook, IL 60062
- Reason
- The product labeling incorrectly lists the particle filtration efficiency as exceeding 99% at 0.1 micron when it should state 95% filtration.
- Distribution pattern
- Nationwide
- Quantity
- 60,000 boxes
- Codes and lots
- all lots of product code 372-8114 (blue) and 372-8125 (pink)
- Recall started
- June 3, 2003
- Device type
- Mask, Surgical
- Root cause (FDA)
- Other
- What the company told customers
- Recalled by letter dated 5/30/03, informing the dentists that the earloop masks were mislabled with the wrong particle filtration efficiency. Correction labels were included with the letter for the dentists to field correct the boxes of masks in their office inventory.
- Recall number
- Z-0944-03