FDA · Medical device recall

Stradis Healthcare Foley Catheter Insertion Tray with preconnected… and 17 more products

Jan. 18, 2023 Class II

Date
Jan. 18, 2023
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Stradis Healthcare
Reason
Potential for the outer bag having pinhole leaks near the seal, which may result in a breach in the sterility of the kit.
Distribution pattern
US Nationwide distribution in the states of FL, IL, IN, MA, MS, NV, OH, TN, TX, WA.

18 products in this recall

Stradis Healthcare Foley Catheter Insertion Tray with preconnected bag and catheter…

Product
Stradis Healthcare Foley Catheter Insertion Tray with preconnected bag and catheter, Sterile
Quantity
920 units
Codes and lots
a. Assembly Number 33066, UDI-DI: M752330661, Lot Numbers: 222782334, 222519530, 222438660; b. Assembly Number 33068, UDI-DI: M752330681, Lot Numbers: 222145412, 220551558, 222508466
Recall started
Dec. 15, 2022
Device type
Urinary Drainage Collection Kit
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0931-2023

PHS Lacertaion Tray…

Product
PHS Lacertaion Tray, Sterile, 20 per case, Manufactured for Professional Hosptial Supply, Inc.
Quantity
300 units
Codes and lots
Assembly Number 477944, UDI-DI: M7524779441, Lot Numbers: 220974270, 220915145
Recall started
Dec. 15, 2022
Device type
Laceration Tray
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0932-2023

Henry Shein CYSTO SET UP PACK…

Product
Henry Shein CYSTO SET UP PACK, 20 per case. Distributed by Henry Schein.
Quantity
1,040 units
Codes and lots
Assembly Number 570-1130, UDI-DI: H65857011301, Lot Numbers: 222990332, 222692396, 222990334, 221013080, 222529321, 220984273, 222692395, 222368551, 220982614, 222992161.
Recall started
Dec. 15, 2022
Device type
Tray, Surgical, Instrument
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0933-2023

Henry Shein GLOVES…

Product
Henry Shein GLOVES, packaged in the following sizes and descriptions: a. Glove 6.5 Pack, 25 Pairs, 10 packs per case; b. Glove 7.0 Pack, 25 Pairs, 10 packs per case; c. Glove 7.5 Pack, 25 Pairs, 10 packs per case; d. Glove 8.0 Pack, 25 Pairs, 10 packs per case; e. Glove 8.5 Pack, 25 Pairs, 10 packs per case; f. Glove 9.0 Pack, 25 Pairs, 10 packs per case;
Quantity
21,755 units
Codes and lots
Assembly Number 570-1130, UDI-DI: H65857011301, Lot Numbers: 222990332, 222692396, 222990334, 221013080, 222529321, 220984273, 222692395, 222368551, 220982614, 222992161. a. Assembly Number 570-1159, UDI-DI: H65857011591, Lot Numbers: 221668682, 222501598, 222308841, 222521158, 222238329, 222276824, 221053079, 221872664; b. Assembly Number 570-1160, UDI-DI: H65857011601, Lot Numbers: 222521156, 222429019, 222428991, 220452660, 222439059, 222428995, 221055116, 222288188, 222501154, 222501155, 222521157, 220762521, 222509319, 222769004, 222639005, 220141429, 222508994; c. Assembly Number 570-1161, UDI-DI: H65857011611, Lot Numbers: 220692475, 222308200, 220251830, 220954105, 220962657, 222248185, 220250698, 222298349, 220251993, 222308198, 222298840, 222521804, 220963077, 220986550, 220141428; d. Assembly Number 570-1162, UDI-DI: H65857011621, Lot Numbers: 221196021, 220462656, 221056018, 221056532, 222308350, 221176016, 221054106, 221056019, 222308552, 222501299, 222501300, 222521301, 220973076, 220896020, 221026017, 213631427; e. Assembly Number 570-1164, UDI-DI: H65857011641, Lot Numbers: 222016037, 222358186, 222430336, 221946035, 222439014, 222430367, 222630366, 222902499, 220252640, 222496579, 221053074, 222350340, 222016033, 222016036, 222146031, 222358129, 222358550, 222496823, 222691515, 220462661, 222692162, 222016034, 222509320, 222692160, 221956029, 220452523, 222016032, 222499000; f. Assembly Number 570-1163, UDI-DI: H65857011631, Lot Numbers: 222290711, 221796024, 221796026, 221796027, 222497161, 222509729, 222510199, 221193075, 222458997, 221024107, 221796025, 222509511, 221796023, 221796022
Recall started
Dec. 15, 2022
Device type
Personal Protection Kit
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0934-2023

Picc Kit…

Product
PICC KIT, 10/CS, ACCESS MEDICAL LABEL
Quantity
180 units
Codes and lots
Assembly Number ACC-2550, UDI-DI: M752ACC25501, Lot Number: 220241506
Recall started
Dec. 15, 2022
Device type
Peripheral Catheter Insertion Kit
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0935-2023

HCT 500mL BAGS…

Product
HCT 500mL BAGS, 2 Port, Sterile, Triple Packaged, 10 EVA BAGS/Pack, 7 Packs PER CASE. Distributed by Health Care Technology.
Quantity
644 units
Codes and lots
Assembly Number HCT10500, UDI-DI: M752HCT105002, Lot Number: 220602531
Recall started
Dec. 15, 2022
Device type
Container, I.V.
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0936-2023

HCT 50mL EVA BAGS…

Product
HCT 50mL EVA BAGS, 2 Port, Sterile, Tripple Packaged, 20 EVA BAGS/Pack, 7 Packs PER CASE. Distributed by Health Care Technology.
Quantity
2,779 units
Codes and lots
Assembly Number HCT2050, UDI-DI: M752HCT20502, Lot Number: 220130152, 221677841
Recall started
Dec. 15, 2022
Device type
Container, I.V.
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0937-2023

HCT 12 CC Syringe Kit…

Product
HCT 12 CC SYRINGE KIT, 20/TRAY,5 TRAYS/BAG, 5 BAGS /CASE HCT. Distributed by Health Care Technology.
Quantity
112,675 units
Codes and lots
Assembly Number HCT3548-2, UDI-DI: M752HCT354821, Lot Numbers: 222770947, 221809694, 221938965, 221748547, 221578546, 221808445, 221386002, 222859338, 222940522, 221938972
Recall started
Dec. 15, 2022
Device type
Syringe, Piston
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0938-2023

HCT 6 CC Syringe Kit…

Product
HCT 6 CC SYRINGE KIT, 25/TRAY, 4 TRAYS/BAG, 5 BAGS/CASE. Distributed by Health Care Technology.
Quantity
76,260 units
Codes and lots
Assembly Number HCT3916-1, UDI-DI: M752HCT391611, Lot Numbers: 222491279, 220113056, 222782510, 221306003, 220051819, 222149971, 222210292, 221245035, 221458554
Recall started
Dec. 15, 2022
Device type
Syringe, Piston
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0939-2023

HCT CENTRAL LINE Insertion Tray…

Product
HCT CENTRAL LINE Insertion Tray, I2 per case. Distributed by Health Care Technology
Quantity
1,176 units
Codes and lots
Assembly Number HCT8249CP-3, UDI-DI: M752HCT8249CP30, Lot Numbers: 222098152, 222551621, 221010141, 222309118, 222901807
Recall started
Dec. 15, 2022
Device type
Central Venous Catheter Tray
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0940-2023

HCT MAX BARRIER PICC INSERTION TRAY 12 per case…

Product
HCT MAX BARRIER PICC INSERTION TRAY 12 per case. Distributed by Health Care Technology
Quantity
200 units
Codes and lots
Assembly Number HCT8428CP-2, UDI-DI: M752HCT8428CP20, Lot Number: 222218373
Recall started
Dec. 15, 2022
Device type
Central Venous Catheter Tray
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0941-2023

Stradis Healthcare NEONATAL PICC Tray *PGBK* 20 trays per case…

Product
Stradis Healthcare NEONATAL PICC Tray *PGBK* 20 trays per case. Manufactured by Stradis Healthcare, Distributed by Med Alliance Group
Quantity
2,500 units
Codes and lots
Assembly Number MAG-3000, UDI-DI: M752MAG30001, Lot Numbers: 220883387, 221226000, 220960490, 220986568, 222074682, 222458932
Recall started
Dec. 15, 2022
Device type
Peripheral Catheter Insertion Kit
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0942-2023

McCormick Medical Distribution PORT TRAY…

Product
McCormick Medical Distribution PORT TRAY, 25 trays per case. Distributed by McCormick Medical Distribution.
Quantity
250 units
Codes and lots
Assembly Number MC-3535PT, UDI-DI: M752MC3535PT1, Lot Numbers: 220673037, 222096813
Recall started
Dec. 15, 2022
Device type
I.V. Start Kit
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0943-2023

Stradis Healthcare TCTR Convenience Kit…

Product
Stradis Healthcare TCTR Convenience Kit. Manufactured by Stradis Healthcare.
Quantity
80 units
Codes and lots
Assembly Number TCTR01, UDI-DI: M752TCTR011, Lot Numbers: 222014334
Recall started
Dec. 15, 2022
Device type
Tray, Surgical, Instrument
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0944-2023

Tacy Medical…

Product
Tacy Medical, Inc Fluid Filter Transfer Kit, Sterile, 60 kits per bag, 5 bags per case. Manufactured for Tracy Medical, Inc.
Quantity
1,175 units
Codes and lots
Assembly Number TM.FFTK, UDI-DI: M752TMFFTK0, Lot Numbers: 220954416, 222418904, 221056552, 220070456, 220400116, 220411513, 220952568, 220943346, 221727862, 222709611, 222701385
Recall started
Dec. 15, 2022
Device type
I.V. Start Kit
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0945-2023

Tacy Medical…

Product
Tacy Medical, Inc Syringe Convenience Pack, IML L/L, Sterile, 60 packs per bag, 5 bags per case. Manufactured for Tracy Medical, Inc.
Quantity
275 units
Codes and lots
Assembly Number TM.SCP1ML, UDI-DI: M752TMSCP1ML0, Lot Numbers: 222270319, 220834335
Recall started
Dec. 15, 2022
Device type
I.V. Start Kit
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0946-2023

Tacy Medical…

Product
Tacy Medical, Inc PORT ACCESS TRAY, Sterile, QTY 20. Manufactured for Tracy Medical, Inc.
Quantity
600 units
Codes and lots
Assembly Number TM8823-1, UDI-DI: M752TM882310, Lot Number: 220110674
Recall started
Dec. 15, 2022
Device type
I.V. Start Kit
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0947-2023

Vital Care Products NEONATAL PICC TRAY…

Product
Vital Care Products NEONATAL PICC TRAY, Sterile, QTY 20. Manufactured for Vital Care Products, Inc.
Quantity
200 units
Codes and lots
Assembly Number VCP16515-1, UDI-DI: M765VCP1651511, Lot Number: 221925083
Recall started
Dec. 15, 2022
Device type
Peripheral Catheter Insertion Kit
Root cause (FDA)
Process control
What the company told customers
Stradis Healthcare notified consignees on about 12/15/2022 via email. The letter instructed consignees to examine inventory, immediately quarantine any affected product, contact Stradis Customer Care to arrange for return shipment of the recalled product, and to notify customers if the product was further distributed. Consignees were also asked to complete and return the Recall Acknowledgement Form.
Recall number
Z-0948-2023

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0931-2023, Z-0932-2023, Z-0933-2023, Z-0934-2023, Z-0935-2023, Z-0936-2023, Z-0937-2023, Z-0938-2023 and others. The agency record governs; check it and the company before acting.