FDA · Medical device recall

VISU 200 on S8 Floor Stand…

Jan. 14, 2011

Date posted by the FDA
Jan. 14, 2011
Status
Terminated
Filed by
Carl Zeiss Meditec Inc
Product
VISU 200 on S8 Floor Stand, Manufactured by Carl Zeiss Surgical GmbH, Germany Medical device: Surgical microscope used during ophthalmic surgeries, which allows real time visualization of the surgical area. Product Quantity Distributed (Int) 2540 units
Reason
Potential Retina Damage-- An error in the user manual may result in misapplied filter and injury during treatment.
Distribution pattern
Worldwide Distribution including USA (nationwide) and the UK
Quantity
2540 units
Codes and lots
User manuals with numbers G-30-1246-e, G-30-1284-e, G-30-1284-en, and G-30-1236-en, affecting all stands.
Recall started
May 18, 2009
Device type
Microscope, Surgical
Root cause (FDA)
Labeling mix-ups
What the company told customers
Carl Zeiss Meditec Inc sent a SAFETY ALERT letter dated May 18, 2009, via Federal Express to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were reminded to use the correct retinal protection filter (GG475) when using the OPMI VISU 200 system during ophthalmic surgery. Verify the position of the retinal protection filter and inform their staff of the customizable feature. Verify that the correct filter was in place. Customers were asked to complete the acknowledgement form indicating receipt of the alert. Sign and return the form to: Carl Zeiss Meditec in Dublin, California via the enclosed self-addressed, pre-paid Federal Express envelope. For any questions customers should call (925) 557-4422.
Recall number
Z-0904-2011