GE Medical Systems Innova 2000 and Innova 2000S… and 2 more products

May 24, 2006

Date posted by the FDA
May 24, 2006
Status
Terminated
Filed by
GE Healthcare
Reason
Incorrect screws were used for mounting the detector which may lead to a possible part falling onto a patient.
Distribution pattern
Nationwide including Puerto Rico and OUS to include: Australia, Belgium, Brazil, Bulgaria, Byelarus, Canada, China, Columbia, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Honduras, Hungary, Indonesia, Iraq, Ireland, Israel, India, Iran, Italy, Jordan, Japan, Republic of Korea, Lebanon, Lithuania, Martinique, Mexico, Morocco, Netherlands, Norway, New Zealand, Poland, Portugal, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Republic of China, Thailand, Tunisia, Turkey, United Kingdom, Venezuela

3 products in this recall

GE Medical Systems Innova 2000 and Innova 2000S…

Product
GE Medical Systems Innova 2000 and Innova 2000S, and Innova 2000 and Innova 2000S mobile version. Cardiovascular Imaging System.
Quantity
856 (534 domestic and 322 OUS)
Codes and lots
all serial numbers
Recall started
March 6, 2006
Device type
Solid State X-Ray Imager (Flat Panel/Digital Imager)
Root cause (FDA)
Other
What the company told customers
GEHC is issuing a validated field action 3/6/2006 to correct the issues. Existing screws/bolts will be replaced using correct torque with screws containing cotter pins. Loctite liquid glue will be applied. Service manuals have been updated with instructions for using correct screws and torque. The field action is being implemented via Field Modification Instruction (FMI) 12033.
Recall number
Z-0896-06

GE Healthcare Innova 3100 and Innova 3100 with Bolus Chasing offered as an option…

Product
GE Healthcare Innova 3100 and Innova 3100 with Bolus Chasing offered as an option, Cardiovascular Imaging System.
Quantity
300 (191 domestic and 109 OUS)
Codes and lots
all devices manufactured before November 2005
Recall started
March 6, 2006
Device type
Solid State X-Ray Imager (Flat Panel/Digital Imager)
Root cause (FDA)
Other
What the company told customers
GEHC is issuing a validated field action 3/6/2006 to correct the issues. Existing screws/bolts will be replaced using correct torque with screws containing cotter pins. Loctite liquid glue will be applied. Service manuals have been updated with instructions for using correct screws and torque. The field action is being implemented via Field Modification Instruction (FMI) 12033.
Recall number
Z-0897-06

GE Healthcare Innova 4100 and Innova 4100 with Bolus Chasing offered as an option…

Product
GE Healthcare Innova 4100 and Innova 4100 with Bolus Chasing offered as an option. Digital Fluoroscopic Imaging System.
Quantity
291 (213 domestic and 78 OUS)
Codes and lots
all devices manufactured before November 2005
Recall started
March 6, 2006
Device type
Solid State X-Ray Imager (Flat Panel/Digital Imager)
Root cause (FDA)
Other
What the company told customers
GEHC is issuing a validated field action 3/6/2006 to correct the issues. Existing screws/bolts will be replaced using correct torque with screws containing cotter pins. Loctite liquid glue will be applied. Service manuals have been updated with instructions for using correct screws and torque. The field action is being implemented via Field Modification Instruction (FMI) 12033.
Recall number
Z-0898-06

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0896-06, Z-0897-06, Z-0898-06. The agency record governs; check it and the company before acting.

Report a mistake on this page