FDA · Medical device recall

SIGNA Premier Nuclear Magnetic Resonance Imaging System… and 3 more products

Feb. 5, 2020 Class II

Date
Feb. 5, 2020
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Distribution pattern
Worldwide distribution - US nationwide distribution and countries of Australia, China, France, Germany, India, Italy, Japan, Netherlands, Oman, Spain, Sweden, Switzerland, Turkey, and United Kingdom.

4 products in this recall

SIGNA Premier Nuclear Magnetic Resonance Imaging System…

Product
SIGNA Premier Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Reason
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
Quantity
62 (30 US, 32 OUS) total devices
Codes and lots
Mfg. Lot or Serial# System ID 33-4-1031 415723SHMR19 To be provided RTD01204 To be provided DPN88672
Recall started
Dec. 23, 2019
Device type
System, Nuclear Magnetic Resonance Imaging
Root cause (FDA)
Software design
What the company told customers
This issue will be resolved by sending an Urgent Medical Device Correction letter to Medical Facility staff (Director of Clinical/Radiology, Risk Manager/Hospital Administrator, Director of Biomedical Engineering) with affected systems. This letter will inform the customer of the issue and instruct the user to ensure the displayed system date is correct. The long-term solution is to correct affected systems with updated software.
Recall number
Z-0874-2020

SIGNA Architect Nuclear Magnetic Resonance Imaging System…

Product
SIGNA Architect Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Reason
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
Quantity
62 (30 US, 32 OUS) total devices
Codes and lots
Mfg. Lot or Serial # System ID PG75A1700023MR 630DMGMR3T 33-4-1252 312702MR3T 033-4-1005 518262MR750W To be provided 082427210212 To be provided 082427170026 To be provided GON4713311 To be provided M4017073 To be provided EM0601 To be provided O009MR03 To be provided GON4634845
Recall started
Dec. 23, 2019
Device type
System, Nuclear Magnetic Resonance Imaging
Root cause (FDA)
Software design
What the company told customers
This issue will be resolved by sending an Urgent Medical Device Correction letter to Medical Facility staff (Director of Clinical/Radiology, Risk Manager/Hospital Administrator, Director of Biomedical Engineering) with affected systems. This letter will inform the customer of the issue and instruct the user to ensure the displayed system date is correct. The long-term solution is to correct affected systems with updated software.
Recall number
Z-0875-2020

SIGNA Pioneer Nuclear Magnetic Resonance Imaging System…

Product
SIGNA Pioneer Nuclear Magnetic Resonance Imaging System - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Reason
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system's date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
Quantity
62 (30 US, 32 OUS) total devices
Codes and lots
Mfg. Lot or Serial # System ID To be provided GON4833100 01264 860536MR3T KZTBL1700064TJ 860871MR3T KZTBL1800069TJ RADNETMR1700 KZTBL1800022TJ 508872PMR 01302 RADNETMR1708 KZTBL1900130TJ 301698MR1 KZTBL1900071TJ 704333PIO KZTBL1900110TJ 631751MR KZTBL1900107TJ RADNETMR1707 KZTBL1900121TJ 803936LMCMR2 033-4-1532 281HMI750W To be provided 83199160995 To be provided M24313812 To be provided M2850013 To be provided TP0013MR01 To be provided PNUE04MR02 To be provided TP0008MR07 To be provided 083027823408018 To be provided 083027823888617 To be provided GON4826677 To be provided GON4762453 To be provided B5142305 To be provided ZC0002 To be provided ZC0034 To be provided ZC0001 To be provided ZC0306 To be provided GON4795546 To be provided 222768MR01 To be provided 34340MRS02 To be provided 10627MRS02
Recall started
Dec. 23, 2019
Device type
System, Nuclear Magnetic Resonance Imaging
Root cause (FDA)
Software design
What the company told customers
This issue will be resolved by sending an Urgent Medical Device Correction letter to Medical Facility staff (Director of Clinical/Radiology, Risk Manager/Hospital Administrator, Director of Biomedical Engineering) with affected systems. This letter will inform the customer of the issue and instruct the user to ensure the displayed system date is correct. The long-term solution is to correct affected systems with updated software.
Recall number
Z-0876-2020

SIGNA Vibrant Nuclear Magnetic Resonance Imaging System…

Product
SIGNA Vibrant Nuclear Magnetic Resonance Imaging System, 1.5T Signa HDx, 1.5T Signa HDxt, Signa HDi - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
Reason
It was identified that due to a potential installation workflow issue, the MR system date could be set incorrectly. The system s date and time setting are used to populate the DICOM header information on images. This could result is an inaccurate date recorded on the images.
Quantity
62 (30 US, 32 OUS) total devices
Codes and lots
Mfg. Lot or Serial # System ID 22D-4-0868 901BMHCRITMR HDGOC1900010SC 209521SGMR HDGOC1900015SC 4577711MR11 002776WHF 954916MR1 022D-4-1175 808243MR 857 LONGWOOD30MR2 00000290815MR0 508580BMR3 002801WHF 734246MR 002930WHF 845279MR 00000269040MR2 412749MR1 022D-4-1125 801RIVMR2 HDGOC1900003SC 206386SMR H05114 4815915MR11 00000277443MR8 304526SMR1 To be provided GON4796599 To be provided 424881MR01 To be provided 083027829345912 To be provided 083027862100408
Recall started
Dec. 23, 2019
Device type
System, Nuclear Magnetic Resonance Imaging
Root cause (FDA)
Software design
What the company told customers
This issue will be resolved by sending an Urgent Medical Device Correction letter to Medical Facility staff (Director of Clinical/Radiology, Risk Manager/Hospital Administrator, Director of Biomedical Engineering) with affected systems. This letter will inform the customer of the issue and instruct the user to ensure the displayed system date is correct. The long-term solution is to correct affected systems with updated software.
Recall number
Z-0877-2020

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0874-2020, Z-0875-2020, Z-0876-2020, Z-0877-2020. The agency record governs; check it and the company before acting.