Signa 1.5T TwinSpeed Magnetic Resonance Imaging System and 3 more products

April 13, 2022 Class II

Date
April 13, 2022
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
GE Healthcare, LLC
Reason
There is potential for the images to be flipped left to right.
Distribution pattern
Worldwide Distribution. US nationwide, Argentina, Australia, Azerbaijan, Bosnia and Herzegovina, Brazil, Canada, China, Denmark, Ecuador, Ethiopia, France, Germany, Greece, Hong Kong, India, Iran, Ireland, Italy, Japan, Korea, Kuwait, Lebanon, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Serbia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syria, Taiwan, Turkey, United Arab Emirates, and United Kingdom.

4 products in this recall

Signa 1.5T TwinSpeed Magnetic Resonance Imaging System

Product
Signa 1.5T TwinSpeed Magnetic Resonance Imaging System
Quantity
10 units
Codes and lots
System IDs: MMIMR01, 213748HDMR, 714125MR, 561218MR1, 918523MR1,304766MR1, 082427030026, 082427080007, 082427020017, 376569401MR
Recall started
Feb. 18, 2022
Device type
System, Nuclear Magnetic Resonance Imaging
Root cause (FDA)
Device Design
What the company told customers
This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions and link for customer to access and download the revised service manual.
Recall number
Z-0857-2022

1.5T SIGNA Excite HD…

Product
1.5T SIGNA Excite HD, 3.0T SIGNA Excite HD. Magnetic Resonance Imaging System
Quantity
76 units
Codes and lots
System IDs: 928223MR3T, 360397MR2, 602406IMR, 480323MR1, 928SMIMR, 3825, 559325CIIMR1, 805988MR2, 6617263TMR, 650306SMR3T, 4036, 3600, 3266, 818RIMR3T, 4267, 352333MR3T, 407SOMR3T, 941PCMR, 386HPOMR, ALLIANCESIG463, 337261MR1, 601288MR3T, 662841MR4, 336713WFUMR, 609383MR, 201487MR, 805998MR1, 805998CR1, 702369SM3TLV, 405634S3T, 814234MR, 6053883T, 731425MR2, 901425MROP1, 806212MR3T, 409892SMR2, 972566MR3, 972378NSC3TMR1, 801377MR2, 910277003, 2577595, 082427090088, 082427210045, 082427270017, 082427160041, 082427040047, 082427040052, 082427040074, 082427210039, 082427090090, 082427310027, 082427100093, 082427140039, 082427030039, 082427100099, 082427100084, 082427070050, 082427070034, 082427310019, 082427100089, IR1503MR01, A5125121, A5806911, YM3034, 850270175, 850270176, 850270199, 850270361, 850270205, KW1004MR03, RF0754MR01, GEE34040, 0853270071, 0853270072, 34208MRS01, 34006MRS03
Recall started
Feb. 18, 2022
Device type
System, Nuclear Magnetic Resonance Imaging
Root cause (FDA)
Device Design
What the company told customers
This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions and link for customer to access and download the revised service manual.
Recall number
Z-0858-2022

SIGNA Excite 3T…

Product
SIGNA Excite 3T. Magnetic Resonance Imaging System
Quantity
55 units
Codes and lots
System IDs: 3422, 209577MRMOBX, 407303CHMR, SIMONVILLAGES3T, 321MR3T3, 321MR3T2, 407999MR3T, 316268TWIN3, 8109873T, 732473MR, 702433SIM3T, 864573MBMR2, 901726OPMR, 214879INVIVO, 817578LGMR2, 972665XDDAMR1 2284SIG15T, 47413, 2550000, 975, 604270MR3T, 416813MR3, 416369MR2, 082427130025 082427130020, 082427130022, 082427270012, 082427270011, 082427150025, 082427020020, 082427010032, 082427160051, 082427160050, 082427120042, 082427120027, 082427030037, 082427250016, 082427070046, 082427150026, 082427090035, 082427140034, 082427140040, 082427090044, 082427100075, 082427120029, 082427070049, 082427070048, 082427220010, 082427300008, YM1408, YM1414, YM1406, 850270152, LB1049MR02, PK1184MR01
Recall started
Feb. 18, 2022
Device type
System, Nuclear Magnetic Resonance Imaging
Root cause (FDA)
Device Design
What the company told customers
This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions and link for customer to access and download the revised service manual.
Recall number
Z-0859-2022

1.5T and 3.0T SIGNA HDx…

Product
1.5T and 3.0T SIGNA HDx, 1.5T and 3.0T SIGNA HDxt, and 1.5T SIGNA HDxt Mobile. Magnetic Resonance Imaging System
Quantity
653 units
Codes and lots
System IDs: 907562P3T, 256734MR3T, 623879MR1, 623433MR1, 480584MR1, 602689MR1, 9287573TMR, 480324MR1, 480614MR1, 928317MR1, 714868MR1, RADNETMR615, 714619MR, 5107923TMR, 858657MR1, 323442MR3, 323442MR1, 213977MR2, 323932MR1, 323442MR2, RADNETMR612, 650259SM3T, RADNETMR1521, RADNETMR592, 626397MR2, RADNETMR620, 916DR3TMR, 415UCLPMR, 415600D3T, 415476UMR6, 415353MRC1, 415MTZMR, 408GS3TMR, RADNETMR1595, 408885MR3T, RADNETMR608, 805569MR, 415723SHMR8, 661AIMR2, RADNETMR594, 805CHLIMR1, 720848MR2, 303340MR, 030022MR01, 515282MR1, 859275MR3T, 2709263HD, 504873TMR3, 504DISWMR, 504DISMR3T, BEV3TMR, 9786823T, RADNETMR1659, RADNETMR1589, 207MMCSMR, 517353VH3, 269342MR3T, 269342MR3T2, 517374MR3T, 248465MR3T, 248849MR3T, 662MSU3T, 406751K3T, CONCORD3T, 201669MR2, 716859MR3T, NYP347652, 212305MR83T, 5857853T, 631686MR3T, 5853413TLAC, 419383MR3T, 503280MR2, 6109883TMR, 423844MR2, 865541MR3T, 615777MR3T, 936441MR1, 817882H3TMR, 513556MR, 713441MR5, 512TSIMR2, 210617TCAMR2, 210617TCAMR1, 801221MR, 509758TMR, 425288CMR, 2063203TMR, 414973MR2, 608785HDMR, 414649MR5, 414973MR1, 422SIGNA3T, 250374MRTS, 604682MRI, 604682MRI2, 604739MR, 082427120201, DK1030MR02, ET1009MR01, 493013MR01, HC1819MR01, 850270170, 850270218, MR281340MR0, DVT0127, PER19516, PER08626, RU1095MR03, ZA2222MR01, MPX82251, GP178328, FPG98423, FPG98415, MEX68900, DPN88605, ME051580, MPX10000, EPP00302, 725030MR01, 00149MRS01, 205939MR4, 205783TSMR, 2059393T, 928213MR3T, 480354MR1, 602588CMR1, 480464MR1, 4809643TMR, 954SIMON3T, 480478MR1, 602406MR6, 831423MR, 4802643TMR, 480425MR5, 480264MR2, 5204493T, RADNETMR590, 310623MR, 626359MR3, 760ENCMR, RADNETMR627, 213777MR, 323268MR2, 213387MR, 510MIA3T, 951SWMR, 805GI3T, 760323MR3T, 650FALK3TMR, 650853PAMR2, RADNETMR617, RADNETMR646, 909806MR, 559585CCICMR2, 916UCDMR3T, 916KM3TMR, RADNETMR609, 858453VMR2, 858822MR3, 4419, RADNETMR1706, 510939JMMR3T, 626962MR, 303SMIMR1, 303633MR, 970244MR3T, 719542MR, 719595MR, 860659MR, 203863MR3T, 860545MR2, 860231MR, 202865MR3T, 407BDIMR, 3173383T1, SIMONMED3TBRANDON, 850916A3T, 407OSCMR3T, 239307MR, NCHMBMR3T, 904276OPTS, PRMTMR3T, 850475IC3T, 727440MR1, 727898MR1, 412692MR1, 813251MR3T, 813972MR3T, 772VRAMR3T, 407SMWPMR1, 7067213TMR, 7705353TMR, 404772NMR3T, 808591MR, 3096653TMR, 773665MR3T, 847SAMC3T, 708383WS3T, 309672MR3T, 859313MR, 225763MR, RADNETMR1636, 301459MR, 410535MR, 410543MR, 3137453TMR, 269341MR3T, 313294MR, 248551MR3T, 2489643T, 9525413TMR, 612626MR3T, 2187493T, 573815BMR1, 636947JMR1, 601268MR3T, 910609CF3T, 402219MR3T, 908725MR, 7326133T, 201632MR, 732632MR3, 609652MR3T, 732223MR, 609497MR3T, 919684MR3, 7024230SM3THL, M48823, RADNETMR554, COR373219, RADNETMR1583, 631982MR, 405CDIMR, 503652MR1, 503216MR3, 503215MR1, 717975MR3, 6053383TMR, 6053223TMR, 731935MR3, 865218MR3T, 901572L3T, 615284MR3T, 615TANMR, 865481MR3T, 865545MR3T, 806NWTH3TMR, 214559SRHMR1, 214329PHIMR1, 214341PDGPMR1, 281333MR1, 713524BIMR1, ACB41121MR, 713704HMR4, 212288GE3T, ACB41109MR, 972665XBPMR1, 210804SUNMR, 801373MR3, 435688MR3, 801321LMR2, 70389MR3T, 703573MR6, 8047643T, 8043303T, 540776MR3, 425297RMR, 304636MR, 304526CMR2, 16683THDX, 2901847MR3T, 910272046, 910273033, 83160776416, AZ4013MR01, BA4080MR01, 2885417, 2854971, MRR8701, 2766544, 2806189, 2871974-1, MRR9017, MRR2396, MRWB0540, 604557MR1, 905522MR, 902678MR1, 416369MR3T, 519254MR, 082427090190, 082427040149, 082427110026, 082427090170, 082427080033, 082427100192, 082427170011, 082427170012, 082427130067, 082427070069, 082427070120, 082427310042, 082427130090, 082427130046, 082427130044, 082427130084, 082427130059, 082427130061, 082427130083, 082427130062, 082427130042, 082427130077, 082427160081, 082427160105, 082427080051, 082427100220, 082427100201, 082427100173, 082427030115, 082427270016, 082427200052, 082427270030, 082427070054, 082427090115, 082427290013, 082427040101, 082427040105, 082427040088, 082427040113, 082427040147, 082427040099, 082427040100, 082427040092, 082427120070, 082427120102, 082427150051, 082427200056, 082427200036, 082427120086, 082427010074, 082427110042, 082427110025, 082427110034, 082427020037, 082427090174, 082427060051, 082427060063, 082427060032, 082427060025, 082427010054, 082427090246, 082427090292, 082427080044, 082427010038, 082427250060, 082427030106, 082427040094, 082427050027, 082427080026, 082427100172, 082427190049, 082427140150, 082427090171, 082427090164, 082427240019, 082427210063, 082427270022, 082427160159, 082427160084, 082427160110, 082427160065, 082427160064, 082427160113, 082427070062, 082427020065, 082427070055, 082427070116, 082427120091, 082427120199, 082427160117, 082427230068, 082427090335, 082427090245, 082427120126, 082427120062, 082427120077, 082427100153, 082427250088, 082427250067, 082427250030, 082427270040, 082427270032, 082427140134, 082427140120, 082427140062, 082427020025, 082427210049, 082427210053, 082427210070, 082427210060, 082427210058, 082427240048, 082427310046, 082427160103, 082427030042, 082427030116, 082427030050, 082427150056, 082427100228, 082427090288, 082427090124, 082427100126, 082427100161, 082427100176, 082427170008, 082427100102, 082427160203, 082427160072, 082427140067, 082427140051, 082427140077, 082427110047, 082427110044, 082427100164, 082427100198, 082427100185, 082427160070, 082427010077, 082427010093, 082427010095, 082427010069, 082427120093, 082427030071, 082427070063, 082427150103, 082427150040, 082427150048, 082427070101, 082427160272, 082427110027, 082427120072, 082427200072, 082427030090, 082427030064, 082427190066, 082427210127, 082427280002, 082427310108, 082427310032, 082427310029, 082427310048, 082427310034, 082427240041, 082427200039, 082427060028, 082427060038, 082427120083, 082427020041, 082427240030, 082427240070, 082427090139, 082427090208, 082427090229, 082427090267, 082427090227, 082427090263, 082427090238, 082427090134, 082427090141, 082427090226, 082427090289, 082427090232, 082427190078, 082427060034, 082427220018, 082427220041, 082427220016, 082427030056, 082427030086, 082427300010, 082427300014, 082427300018, 082427300012, 082427020026, 082427190075, 082427190069, 082427100119, 082427100309, 082427100182, 082427090173, 082427010141, 082427010047, 082427010150, 082427240034, 082427010065, DK1018MR01, ET1293MR02, S41600123, HC0074MR01, MC6946MR07, H10392MR03, H2193MR05, 0828270039, 083027828659711, 083027041002010, 083027601000510, 083027609389015, 083027605826311, 083027202000249, 083027024220111, 083027022101691, 083027202001698, 083027202101396, 083027207848338, 083027002101567, 083027006495511, 083027002101706, 083027009445910, 083027007755738, 083027002730811, 083027202101819, 083027243783711, 083027862100837, IE1111MR01, A469231301, A5632416, C5323526, YM1404, YM1261, YM1799, YM0867, YM3040, YM0873, YM1836, YM1839, YM2021, YM1851, YM1837, YM1831, YM1817, YM1832, YM1821, YM1401, YM3900, YM1805, YM1395, YM3883, YM1815, YM1841, YM3032, YM1797, YM1794, YM1881, YM1796, YM1788, YM1842, YM1834, YM1840, YM1800, YM1882, YM1813, YM1812, YM1850, YM1823, YM1843, YM1810, YM1845, YM1820, YM2018, YM1400, YM0796, YM0790, YM1870, YM0888, YM3042, YM1804, YM1880, YM1877, YM2003, YM1808, YM0802, YM1411, YM1798, YM1819, YM1809, YM1822, YM0801, YM1848, YM1828, YM1844, YM1792, YM1793, YM1402, YM1818, YM2020, YM1824, YM1846, YM0791, YM1879, YM1399, YM1826, YM1849, YM1847, YM3939, YM0872, YM2005, YM1827, YM1803, YM2012, YM1835, YM1816, 850270200, 850270202, 850270333, 850270220, LB1000MR02, 2716757SIGNAHX, 920279024, PL3800MR01, PL3545MR01, PL1050MR01, PL3040MR01, PL3546MR01, PER28007, 787641HDXT, RO1189MR01, RU1588MR01, RU1087MR01, RU1089MR04, RU1067MR02, RU1688MR01, RU9155MR02, RU1884MR01, SA1012MR01, SA1001MR03, RS4202MR01, SI4024MR02, ZA2169MR01, DEL26444, CEN03406, DPC79014, DEN51500, NPV84502, G002MR01, SY2012MR02, 0853270074, 0853270095, 0853270081, 0853270094, 0853270088, AE1190MR01, 70328MRS04
Recall started
Feb. 18, 2022
Device type
System, Nuclear Magnetic Resonance Imaging
Root cause (FDA)
Device Design
What the company told customers
This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected systems. This 2/18/2022 letter informs the customer of the issue and provides actions to be taken for the continued use of the device. GE Healthcare will provide a revised service manual that includes a required Finalization step that ensures a geometry check with a DQA phantom (or similar) following Gradient Switch servicing to ensure cable placement is correct and will have no impact on image orientation. A second letter was send to customers on 07/27/2022 and included instructions and link for customer to access and download the revised service manual.
Recall number
Z-0860-2022

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0857-2022, Z-0858-2022, Z-0859-2022, Z-0860-2022. The agency record governs; check it and the company before acting.