FDA · Medical device recall
VITROS CHEMISTRY PRODUCTS TRIG Slides…
- Date
- Dec. 31, 2014
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- Ortho-Clinical Diagnostics
- Product
- VITROS CHEMISTRY PRODUCTS TRIG Slides, REF 133 6544, IVD For in vitro diagnostic use only. VITROS Chemistry Products TRIG Slides quantitatively measure triglyceride (TRIG) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
- Reason
- Potential for biased results using the specific lot of VITROS CHEMISTRY PRODUCTS TRIG slides.
- Distribution pattern
- Worldwide Distribution - USA (nationwide) Puerto Rico and Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
- Quantity
- USA: 380 units, Foreign: 155 units
- Codes and lots
- Lot 0704-3318-4774 (expiry date 01 Dec 2015)
- Recall started
- Nov. 6, 2014
- Device type
- Lipase Hydrolysis/Glycerol Kinase Enzyme, Triglycerides
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Ortho Clinical Diagnostics sent an Urgent Product Correction Notification dated November 6, 2014, via FedEx overnight courierand/or ORTHO PLUS e-Communications and/or US Postal Service Priority Mail (for PO Boxes only) to all affected customers. The letter identified the product, the problem, and the acton to be taken by the customer. The letter informed the customers of the issue and advise them to discontinue use of the affected lot and to inform their Laboratory Medical Director if they suspect that a previously reported result may have been affected. For additional questions, please contact OCD's Customer Technical Services at 1-800-421-3311. For questions regarding this recall call 908-218-8776.
- Recall number
- Z-0848-2015
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0848-2015. The agency record governs; check it and the company before acting.