Menu
FDA · Medical device recall

ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST)…

Feb. 7, 2024 Class II

Date
Feb. 7, 2024
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Ethicon Endo-Surgery Inc
Product
ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic Linear Cutter Reload (+Gripping Surface technology), 45MM WHITE VASCULAR/THIN, Product No. GST45W
Reason
There is the potential for incomplete staple line formation during the firing sequence, without most or all staples, which may lead to insufficient tissue apposition that could require surgical intervention. This may result in bleeding, prolonged surgery, or surgical delay.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Canada and Brazil.
Quantity
22,616 (US), 4,476 (OUS)
Codes and lots
UDI-DI 10705036014928 Lots 442C02 and 442C03
Recall started
Dec. 11, 2023
Device type
Staple, Implantable
Root cause (FDA)
Under Investigation by firm
What the company told customers
On December 11, 2023, the firm notified affected consignees through URGENT VOLUNTARY MEDICAL DEVICE RECALL (REMOVAL) letters. Customers were instructed to cease use and distribution of affected product lots. Customers should examine their inventory immediately to determine if they have affected product. Affected product should be quarantined. Unused affected product should be returned. If you require any assistance with returning product, please contact Sedgwick at 888-591-5087 and reference Event # 6925. If you have additional questions regarding this voluntary medical device recall, please contact the Ethicon Resource Department at 1-877-ETHICON (1-877-384-4266). The Ethicon Resource Department is open Monday through Friday, 8:00 AM to 5:00 PM ET.