FDA · Medical device recall
ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST)…
- Date
- Feb. 7, 2024
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- Ethicon Endo-Surgery Inc
- Product
- ENDOPATH ECHELON 45mm RELOADS WITH GRIPPING SURFACE TECHNOLOGY (GST), ECHELON Endopath Stapler, Endoscopic Linear Cutter Reload (+Gripping Surface technology), 45MM WHITE VASCULAR/THIN, Product No. GST45W
- Reason
- There is the potential for incomplete staple line formation during the firing sequence, without most or all staples, which may lead to insufficient tissue apposition that could require surgical intervention. This may result in bleeding, prolonged surgery, or surgical delay.
- Distribution pattern
- Worldwide distribution - US Nationwide and the countries of Canada and Brazil.
- Quantity
- 22,616 (US), 4,476 (OUS)
- Codes and lots
- UDI-DI 10705036014928 Lots 442C02 and 442C03
- Recall started
- Dec. 11, 2023
- Device type
- Staple, Implantable
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- On December 11, 2023, the firm notified affected consignees through URGENT VOLUNTARY MEDICAL DEVICE RECALL (REMOVAL) letters. Customers were instructed to cease use and distribution of affected product lots. Customers should examine their inventory immediately to determine if they have affected product. Affected product should be quarantined. Unused affected product should be returned. If you require any assistance with returning product, please contact Sedgwick at 888-591-5087 and reference Event # 6925. If you have additional questions regarding this voluntary medical device recall, please contact the Ethicon Resource Department at 1-877-ETHICON (1-877-384-4266). The Ethicon Resource Department is open Monday through Friday, 8:00 AM to 5:00 PM ET.
- Recall number
- Z-0834-2024
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0834-2024. The agency record governs; check it and the company before acting.