21Y14 PushValve Product Usage: The PushValves/ MagValves is intended… and 2 more products

Feb. 20, 2019 Class II

Date
Feb. 20, 2019
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Ottobock Orthopedic Industrie
Reason
Some of the installed valve-inserts have a diameter smaller than the specification.
Distribution pattern
US Nationwide Distribution

3 products in this recall

21Y14 PushValve Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting…

Product
21Y14 PushValve Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting of lower limb amputees. The valve is a threadless one-way valve. It is used to evacuate air from the prosthetic socket.
Quantity
538 units
Codes and lots
The products with an article number listed below and delivered between May 1, 2017 and January 19, 2018
Recall started
Jan. 10, 2019
Device type
Joint, Knee, External Limb Component
Root cause (FDA)
Process control
What the company told customers
The firm initiated the recall by email on 01/10/2019 and followed with a letter on 01/11/2019. The letter identified the affected product, problem, actions to be taken and that a second letter with the replacement unit was forthcoming.
Recall number
Z-0826-2019

21Y14=S PushValve upper part Product Usage: The PushValves/ MagValves is intended exclusively for the…

Product
21Y14=S PushValve upper part Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting of lower limb amputees. The valve is a threadless one-way valve. It is used to evacuate air from the prosthetic socket.
Quantity
1 unit
Codes and lots
The products with an article number listed below and delivered between May 1, 2017 and January 19, 2018
Recall started
Jan. 10, 2019
Device type
Joint, Knee, External Limb Component
Root cause (FDA)
Process control
What the company told customers
The firm initiated the recall by email on 01/10/2019 and followed with a letter on 01/11/2019. The letter identified the affected product, problem, actions to be taken and that a second letter with the replacement unit was forthcoming.
Recall number
Z-0827-2019

21Y15 MagValve Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting of…

Product
21Y15 MagValve Product Usage: The PushValves/ MagValves is intended exclusively for the prosthetic fitting of lower limb amputees. The valve is a threadless one-way valve. It is used to evacuate air from the prosthetic socket.
Quantity
110 units
Codes and lots
The products with an article number listed below and delivered between May 1, 2017 and January 19, 2018
Recall started
Jan. 10, 2019
Device type
Joint, Knee, External Limb Component
Root cause (FDA)
Process control
What the company told customers
The firm initiated the recall by email on 01/10/2019 and followed with a letter on 01/11/2019. The letter identified the affected product, problem, actions to be taken and that a second letter with the replacement unit was forthcoming.
Recall number
Z-0828-2019

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0826-2019, Z-0827-2019, Z-0828-2019. The agency record governs; check it and the company before acting.