FDA · Medical device recall
Centricity Universal Viewer 6.0…
- Date
- Jan. 29, 2020
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- GE Healthcare, LLC
- Product
- Centricity Universal Viewer 6.0, a device that displays medical images and data from various imaging sources, and from other healthcare information sources.
- Reason
- Centricity Universal Viewer measurements saved into a DICOM Grayscale Presentation State are incorrect in subsequent views for exams containing series with different pixel sizes and may lead to a potential misdiagnosis.
- Distribution pattern
- Distributed nationwide; Internationally distributed to Mexico and 24 additional countries.
- Quantity
- 205
- Codes and lots
- Model 2088026-9XX Versions 6.0 SP10 or higher
- Recall started
- Dec. 16, 2019
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design
- What the company told customers
- The firm initiated their correction on 12/16/2019 by letter. The letter explained the issue and directed the consignee to follow instructions in the letter until an update to the Centricity Universal Viewer can be made by the firm to correct the issue.
- Recall number
- Z-0819-2020